Abstract

PURPOSE: Open reduction and internal fixation of radial neck fractures can lead to secondary loss of reduction and nonunion due to insufficient stability. Nevertheless, there are only a few biomechanical studies about the stability achieved by different osteosynthesis constructs.
METHODS: Forty-eight formalin-fixed, human proximal radii were divided into 6 groups according to their bone density (measured by dual-energy x-ray absorptiometry). A 2.7-mm gap osteotomy was performed to simulate an unstable radial neck fracture, which was fixed with 3 nonlocking implants: a 2.4-mm T plate, a 2.4-mm blade plate, and 2.0-mm crossed screws, and 3 locking plates: a 2.0-mm LCP T plate, a 2.0-mm 6x2 grid plate, and a 2.0-mm radial head plate. Implants were tested under axial (N/mm) and torsional (Ncm/ degrees ) loads with a servohydraulic materials testing machine.
RESULTS: The radial head plate was significantly stiffer than all other implants under axial as well as under torsional loads, with values of 36 N/mm and 13 Ncm/ degrees . The second-stiffest implant was the blade plate, with values of 20 N/mm and 6 Ncm/ degrees . The weakest implants were the 2.0-mm LCP, with values of 6 N/mm and 2 Ncm/ degrees , and the 2.0-mm crossed screws, with values of 18 N/mm and 2 Ncm/ degrees . The 2.4-mm T plate, with values of 14 N/mm and 4 Ncm/ degrees , and the 2.0-mm grid plate, with values of 8 N/mm and 4 Ncm/ degrees came to lie in the midfield.
CONCLUSIONS: The 2.0-mm angle-stable plates-depending on their design-allow fixation with comparable or even higher stability than the bulky 2.4-mm nonlocking implants and 2.0-mm crossed screws.

Entities:  

Mesh:

Year:  2007        PMID: 18070646     DOI: 10.1016/j.jhsa.2007.08.023

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  16 in total

Review 1.  [Fractures of the elbow joint].

Authors:  T E Nowak; S O Dietz; K J Burkhart; L P Müller; P M Rommens
Journal:  Chirurg       Date:  2012-02       Impact factor: 0.955

2.  [Endoprosthetics of acute radial head fractures].

Authors:  K J Burkhart; K Wegmann; C Ries; L P Müller
Journal:  Unfallchirurg       Date:  2013-08       Impact factor: 1.000

Review 3.  [Mason I fracture - a simple injury?].

Authors:  K J Burkhart; S Franke; K Wegmann; C Ries; F Dehlinger; L P Müller; B Hollinger
Journal:  Unfallchirurg       Date:  2015-01       Impact factor: 1.000

Review 4.  Radial head fractures.

Authors:  R P van Riet; Mpj van den Bekerom; A Van Tongel; C Spross; R Barco; A C Watts
Journal:  Shoulder Elbow       Date:  2019-09-25

5.  [Locking plate osteosynthesis of the radial head fractures : clinical and radiological results].

Authors:  K J Burkhart; D Gruszka; S Frohn; K Wegmann; P M Rommens; C M Eicker; L P Müller
Journal:  Unfallchirurg       Date:  2015-11       Impact factor: 1.000

6.  The role of locking technology in the upper extremity.

Authors:  Siddharth B Joglekar; Asif M Ilyas
Journal:  J Hand Microsurg       Date:  2010-01-08

7.  Treatment of radial head and neck fractures: in favor of anatomical reconstruction.

Authors:  K J Burkhart; K Wegmann; J Dargel; C Ries; L P Mueller
Journal:  Eur J Trauma Emerg Surg       Date:  2012-09-13       Impact factor: 3.693

Review 8.  Biomechanical Concepts for Fracture Fixation.

Authors:  Michael Bottlang; Christine E Schemitsch; Aaron Nauth; Milton Routt; Kenneth A Egol; Gillian E Cook; Emil H Schemitsch
Journal:  J Orthop Trauma       Date:  2015-12       Impact factor: 2.512

9.  Functional Outcome of Resection versus Reconstruction in Mason II-III Radial Head Fractures; A Short-Term Prospective Study.

Authors:  Pradeep Kumar Meena; Sahil Gaba; Sandesh Bobade; Rajendra Verma; Amrut Borade; Jayesh Sonaje; Ankit Chouhan
Journal:  Bull Emerg Trauma       Date:  2017-10

10.  Options and limits of angle stable plates in the treatment of comminuted radial head fractures.

Authors:  T F Raven; L Banken; J Doll; F Westhauser; B Reible; M Schönewald; G Schmidmaier; A Moghaddam
Journal:  J Orthop       Date:  2018-09-06