B A Paldus, C C Harb, T G Spence, R N Zare, C Gmachl, F Capasso, D L Sivco, J N Baillargeon, A L Hutchinson, A Y Cho.
Abstract
Cavity ringdown spectra of ammonia at 10 parts in 10(9) by volume (ppbv) and higher concentrations were recorded by use of a 16-mW continuous-wave quantum-casacde distributed-feedback laser at 8.5 mum whose wavelength was continuously temperature tuned over 15 nm. A sensitivity (noise-equivalent absorbance) of 3.4x10(-9) cm(-1) Hz(-1/2) was achieved for ammonia in nitrogen at standard temperature and pressure, which corresponds to a detection limit of 0.25 ppbv.Entities:
Year: 2000 PMID: 18064145 DOI: 10.1364/ol.25.000666