Experimental investigations for mechanical joint strength following ultrasonically welded pin osteosynthesis.

Heike Meissner1, Eckart Pilling, Gert Richter, Rainer Koch, Uwe Eckelt, Bernd Reitemeier.   

Abstract

The purpose of this study was to determine whether fixation of cranial bone segments using ultrasonically welded pin osteosynthesis showed differences in mechanical stability as compared to fixation of cranial bone segments using screw osteosynthesis. Right and left cranial bone segments from each of 16 young sheep were obtained by craniotomy and re-fixed: on the right with a mesh plate and pins, and on the left with a mesh plate and screws. All osteosynthesis materials consisted of PDLLA, fully amorphous polyactid. A total of 167 cranial bone / mesh plate segments from 16 animals were investigated; 84 segments were pin-fixed and 83 segments were screw-fixed. The implantation time of the re-fixed segments ranged from 1 day to 196 days. The mechanical methods chosen for simulation of stress on the bone segment bonds were two bending tests (horizontal and vertical directions) and a tensile test. The values obtained in the mechanical tests indicate differences in the bond strength between the pin- and screw- fixation methods over the length of in vivo implantation time. The mechanical stability of the ultrasonically welded pin osteosynthesis bonds over the screw osteosynthesis bonds proved to be statistically significant. The implication of these findings should also be relevant in the field of medicine.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18060480     DOI: 10.1007/s10856-007-3074-x

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  7 in total

1.  [Biodegradable miniplates (lactosorb) in cranio-osteoplasty--experimental results with the rapidly maturing, juvenile minipig].

Authors:  H J Becker; J Wiltfang; H A Merten; H G Luhr
Journal:  Mund Kiefer Gesichtschir       Date:  1999-09

2.  Bioabsorbable thread for tight tying of bones.

Authors:  K Hattori; N Tomita; S Tamai; Y Ikada
Journal:  J Orthop Sci       Date:  2000       Impact factor: 1.601

3.  Study by finite element method of the mechanical stress of selected biodegradable osteosynthesis screws in sagittal ramus osteotomy.

Authors:  P Maurer; S Holweg; W-D Knoll; J Schubert
Journal:  Br J Oral Maxillofac Surg       Date:  2002-02       Impact factor: 1.651

Review 4.  [Absorbable and bioconvertible osteosynthesis materials in maxillofacial surgery].

Authors:  J A Obwegeser
Journal:  Mund Kiefer Gesichtschir       Date:  1998-11

5.  New bioresorbable pin for the reduction of small bony fragments: design, mechanical properties and in vitro degradation.

Authors:  L E Claes; A A Ignatius; K E Rehm; C Scholz
Journal:  Biomaterials       Date:  1996-08       Impact factor: 12.479

6.  Fixation of mandibular fractures with biodegradable plates and screws.

Authors:  Kaan C Yerit; Georg Enislidis; Christian Schopper; Dritan Turhani; Felix Wanschitz; Arne Wagner; Franz Watzinger; Rolf Ewers
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2002-09

7.  Resorbable plate osteosynthesis of dislocated or pathological mandibular fractures: a prospective clinical trial of two amorphous L-/DL-lactide copolymer 2-mm miniplate systems.

Authors:  Constantin A Landes; Susanne Kriener; Michael Menzer; Adorjàn F Kovàcs
Journal:  Plast Reconstr Surg       Date:  2003-02       Impact factor: 4.730

  7 in total
  2 in total

1.  The use of BoneWelding® technology in spinal surgery: an experimental study in sheep.

Authors:  Dorothee Heidenreich; Jens D Langhoff; Katja Nuss; Katharina Kluge; Käthi Kämpf; Katalin Zlinsky; Monika Hilbe; Jörg Mayer; Brigitte von Rechenberg
Journal:  Eur Spine J       Date:  2011-04-27       Impact factor: 3.134

2.  Treatment of calvarial defects by resorbable and non-resorbable sonic activated polymer pins and mouldable titanium mesh in two dogs: a case report.

Authors:  Pierre Langer; Cameron Black; Padraig Egan; Noel Fitzpatrick
Journal:  BMC Vet Res       Date:  2018-06-22       Impact factor: 2.741

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.