Literature DB >> 18056531

Direct left atrial pressure monitoring in ambulatory heart failure patients: initial experience with a new permanent implantable device.

Jay Ritzema1, Iain C Melton, A Mark Richards, Ian G Crozier, Chris Frampton, Robert N Doughty, James Whiting, Saibal Kar, Neal Eigler, Henry Krum, William T Abraham, Richard W Troughton.   

Abstract

BACKGROUND: We describe the first human experience with a permanently implantable, direct left atrial pressure (LAP) monitoring system in ambulatory patients with chronic heart failure. METHODS AND
RESULTS: Eight patients with established heart failure and at least 1 heart failure hospitalization or unplanned visit for parenteral therapy in the last year underwent device implantation under fluoroscopic guidance. All subjects received aspirin 150 mg and clopidogrel 75 mg daily. Subjects measured LAP twice daily and attended a clinic regularly for data upload and device calibration. Right heart catheterization was performed at the time of device implantation and at 12 weeks. The device was implanted in all subjects with no procedural complications. At the 12-week follow-up, 87% of device LAP measurements were within +/-5 mm Hg of simultaneous pulmonary capillary wedge pressure readings over a wide range of pressures (1.6 to 71 mm Hg). Net drift corrected by calibration was -0.2+/-1.9 mm Hg/mo. During short-term follow-up, there were no device-related complications or systemic emboli. There were no deaths, no unplanned heart failure clinic visits, and no admissions for heart failure.
CONCLUSIONS: Ambulatory monitoring of direct LAP with a new implantable device was well tolerated, feasible, and accurate at a short-term follow-up. Further follow-up and investigation are warranted to evaluate the clinical utility of LAP monitoring in patients with heart failure.

Entities:  

Mesh:

Year:  2007        PMID: 18056531     DOI: 10.1161/CIRCULATIONAHA.107.702191

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  39 in total

1.  ISHNE/EHRA expert consensus on remote monitoring of cardiovascular implantable electronic devices (CIEDs).

Authors:  Sergio Dubner; Angelo Auricchio; Jonathan S Steinberg; Panos Vardas; Peter Stone; Josep Brugada; Ryszard Piotrowicz; David L Hayes; Paulus Kirchhof; Günter Breithardt; Wojciech Zareba; Claudio Schuger; Mehmet K Aktas; Michal Chudzik; Suneet Mittal; Niraj Varma
Journal:  Ann Noninvasive Electrocardiol       Date:  2012-01       Impact factor: 1.468

2.  [Diminishing borders between cardiology and cardiothoracic surgery: quo vadis?].

Authors:  T Schilling; R Bekeredjian; A Haverich; H A Katus
Journal:  Chirurg       Date:  2010-12       Impact factor: 0.955

Review 3.  Remote hemodynamic monitoring for ambulatory left ventricular assist device patients.

Authors:  Brent C Lampert; Sitaramesh Emani
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

Review 4.  Microelectromechanical systems and nephrology: the next frontier in renal replacement technology.

Authors:  Steven Kim; Shuvo Roy
Journal:  Adv Chronic Kidney Dis       Date:  2013-11       Impact factor: 3.620

Review 5.  Remote Monitoring to Reduce Heart Failure Readmissions.

Authors:  Sitaramesh Emani
Journal:  Curr Heart Fail Rep       Date:  2017-02

Review 6.  Pathophysiology of the transition from chronic compensated and acute decompensated heart failure: new insights from continuous monitoring devices.

Authors:  Philip B Adamson
Journal:  Curr Heart Fail Rep       Date:  2009-12

7.  Entrapped left atrial pressure monitoring catheter in a prosthetic mitral valve.

Authors:  Sathappan Karuppiah; Gladdy George; Binila Chacko; Madhu Andrew Philip
Journal:  Ann Card Anaesth       Date:  2015 Apr-Jun

8.  Ambulatory monitoring of congestive heart failure by multiple bioelectric impedance vectors.

Authors:  Dirar S Khoury; Mihir Naware; Jeff Siou; Andreas Blomqvist; Nilesh S Mathuria; Jianwen Wang; Hue-Teh Shih; Sherif F Nagueh; Dorin Panescu
Journal:  J Am Coll Cardiol       Date:  2009-03-24       Impact factor: 24.094

9.  Therapeutic implications of implantable device-based monitoring of patients with heart failure.

Authors:  Eric Popjes; John P Boehmer
Journal:  Curr Treat Options Cardiovasc Med       Date:  2008-09

Review 10.  Implantable devices to monitor patients with heart failure.

Authors:  Nikolaos Karamichalakis; John Parissis; George Bakosis; Vasiliki Bistola; Ignatios Ikonomidis; Antonios Sideris; Gerasimos Filippatos
Journal:  Heart Fail Rev       Date:  2018-11       Impact factor: 4.214