The effect of arthrodesis, implant stiffness, and time on the canine lumbar spine.

Marc A Asher1, William L Carson, James W Hardacker, Richard G Lark, Sue Min Lai.   

Abstract

STUDY
DESIGN: Canine posterior lumbar instrumentation and fusion.
OBJECTIVES: To study effects of implant rod size and time on the stiffness of related spine construct elements. SUMMARY OF BACKGROUND DATA: The ideal stiffness of posterior spinal implants to successfully treat clinical instability or deformity with minimal side effects is unknown.
METHODS: Twenty-six canines were divided into 7 groups: control, and 6 or 12-month survival after sham or lumbar L3-5 arthrodesis (facet, posterior, and posterolateral) with either 4.76 or 6.35 mm diameter rod-pedicle screw instrumentation. Axial flexion-compression stiffness of the L3-5 segment components and axial compression stiffness of the bypassed and adjacent anterior column elements were measured.
RESULTS: Posterior instrumentation initially increased flexion-compression stiffness of the L3-5 segment more than the intrinsic stiffness of the implant due to control of spinal column flexion buckling. Sham operation did likewise, apparently by posterior scar tissue tethering. The percent contribution of the implant construct to instrumented segment stiffness was significantly less at 6 months without further change from 6 to 12 months; 14% and 22% for 4.76 and 6.35 mm rod constructs, respectively. Spinal column as well as posterior column stiffness after fusion was independent of rod size at 6 months and increased at 12 months in only the 4.76 mm rod group. Bypassed L4 vertebral body stiffness decreased significantly at 6 months, was not rod size dependent and changed little between 6 and 12 months. Bypassed disk stiffness responded in a biphasic manner, apparently increasing at 6 months with significant decrease from 6 to 12 months. Adjacent disk compression stiffness progressively decreased over time independent of rod size, also decreasing after sham operation.
CONCLUSIONS: Both rod sizes were associated with 100% fusion and produced similar changes in bypassed bone and disks, and adjacent disks. There was delayed fusion stress shielding by 6.35 mm rod constructs.

Entities:  

Mesh:

Year:  2007        PMID: 18046166     DOI: 10.1097/BSD.0b013e31804c98e5

Source DB:  PubMed          Journal:  J Spinal Disord Tech        ISSN: 1536-0652


  4 in total

1.  Posterior lumbar fusion by peek rods in degenerative spine: preliminary report on 30 cases.

Authors:  F De Iure; G Bosco; M Cappuccio; S Paderni; L Amendola
Journal:  Eur Spine J       Date:  2012-03-09       Impact factor: 3.134

2.  Flexible Stabilisation of the Degenerative Lumbar Spine Using PEEK Rods.

Authors:  Jacques Benezech; Bruno Garlenq; Gilles Larroque
Journal:  Adv Orthop       Date:  2016-02-15

3.  A study to compare the efficacy of polyether ether ketone rod device with titanium devices in posterior spinal fusion in a canine model.

Authors:  Nanxiang Wang; Huanxin Xie; Chunyang Xi; Han Zhang; Jinglong Yan
Journal:  J Orthop Surg Res       Date:  2017-03-09       Impact factor: 2.359

4.  Biomaterials in Spinal Implants: A Review.

Authors:  Andrew Warburton; Steven J Girdler; Christopher M Mikhail; Amy Ahn; Samuel K Cho
Journal:  Neurospine       Date:  2019-11-04
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.