Abstract

OBJECTIVE: The aim of this intravascular ultrasound (IVUS) study was to assess the efficacy of the AXXESS Plus stent system for the treatment of bifurcation coronary lesions.
BACKGROUND: The AXXESS Plus is a novel bifurcation drug-eluting stent, comprised of a self-expanding flare-shaped stent platform and bioabsorbable polymer coating that releases Biolimus A9.
METHODS: Data were obtained from the AXXESS PLUS trial, a prospective, multicenter, nonrandomized, single-arm study to evaluate safety and efficacy. Six-month follow-up IVUS analysis was available in 49 cases. Volumetric analysis using Simpson's method within the AXXESS stent, and cross-sectional analysis at the ostium of main branch and/or side branch was performed. Impact of bifurcation angle on stent expansion at the carina was also evaluated.
RESULTS: Within the AXXESS stent, neointimal volume obstruction percentage was 2.3% +/- 2.2%, with a minimum lumen area of 7.9 +/- 2.6 mm(2). Lumen area was 5.2 +/- 1.7 mm(2) at main branch ostium, and 4.0 +/- 1.5 mm(2) at side branch ostium. In two cases, incomplete stent apposition was observed at the proximal edge of the AXXESS stent. In one case, a gap between the AXXESS stent and an additional stent was observed. Greater bifurcation angle inversely correlated with smaller stent area at side branch ostium (r = -0.54, P = 0.03) but not at main branch ostium (r = -0.2, P = 0.29).
CONCLUSIONS: This novel self-expanding, drug-eluting bifurcation stent demonstrated effective lesion coverage along with significant neointimal suppression equivalent to current generation balloon-expandable drug-eluting stent technology. Copyright 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18044777     DOI: 10.1002/ccd.21269

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  2 in total

Review 1.  Applications of grayscale and radiofrequency intravascular ultrasound to image atherosclerotic plaque.

Authors:  Somjot S Brar; Gary S Mintz; Akiko Maehara; Gregg W Stone
Journal:  J Nucl Cardiol       Date:  2010-10       Impact factor: 5.952

2.  Expansion of the Multi-Link Frontier™ coronary bifurcation stent: micro-computed tomographic assessment in human autopsy and porcine heart samples.

Authors:  Stefan Kralev; Benjamin Haag; Jens Spannenberger; Siegfried Lang; Marc A Brockmann; Soenke Bartling; Alexander Marx; Karl-Konstantin Haase; Martin Borggrefe; Tim Süselbeck
Journal:  PLoS One       Date:  2011-07-21       Impact factor: 3.240