| Literature DB >> 18044623 |
Kensaku Mori1, Daisuke Deguchi, Kazuyoshi Ishitani, Takayuki Kitasaka, Yasuhito Suenaga, Yosihnori Hasegawa, Kazuyoshi Imaizumi, Hirotsugu Takabatake.
Abstract
This paper presents a method for bronchoscope tracking without any fiducial markers using an ultra-tiny electromagnetic tracker (UEMT) for a bronchoscopy guidance system. The proposed method calculates the transformation matrix, which shows the relationship between the coordinates systems of the pre-operative CT images and the UEMT, by registering bronchial branches segmented from CT images and points measured by the UEMT attached at the tip of a bronchoscope. We dynamically compute the transformation matrix for every pre-defined number of measurements. We applied the proposed method to a bronchial phantom in several experimental environments. The experimental results showed the proposed method can track a bronchoscope camera with about 3.3mm of target registration error (TRE) for wood table environment and 4.0mm of TRE for examination table environment.Mesh:
Year: 2007 PMID: 18044623 DOI: 10.1007/978-3-540-75759-7_78
Source DB: PubMed Journal: Med Image Comput Comput Assist Interv