MOSFET detectors in quality assurance of tomotherapy treatments.

Amanda Cherpak1, Ryan C N Studinski, Joanna E Cygler.   

Abstract

BACKGROUND AND
PURPOSE: The purpose of this work was to characterize metal oxide semiconductor field-effect transistors (MOSFETs) in a 6 MV conventional linac and investigate their use for quality assurance of radiotherapy treatments with a tomotherapy Hi-Art unit.
MATERIALS AND METHODS: High sensitivity and standard sensitivity MOSFETs were first calibrated and then tested for reproducibility, field size dependence, and accuracy of measuring surface dose in a 6 MV beam as well as in a tomotherapy Hi-Art unit. In vivo measurements were performed on both a RANDO phantom and several head and neck cancer patients treated with tomotherapy and compared to TLD measurements and treatment plan doses to evaluate the performance of MOSFETs in a high gradient radiation field.
RESULTS: The average calibration factor found was 0.345+/-2.5%cGy/mV for the high sensitivity MOSFETs tested and 0.901+/-2.4%cGy/mV for the standard sensitivity MOSFETs. MOSFET measured surface doses had an average agreement with ion chamber measurements of 1.55% for the high sensitivity MOSFET and 5.23% for the standard sensitivity MOSFET when averaged over all trials and field sizes tested. No significant dependence on field size was found for the standard sensitivity MOSFETs, however a maximum difference of 5.34% was found for the high sensitivity MOSFET calibration factors in the field sizes tested. Measurements made with MOSFETS on head and neck patients treated on a tomotherapy Hi-Art unit had an average agreement of (3.26+/-0.03)% with TLD measurements, however the average of the absolute difference between the MOSFET measurements and the treatment plan skin doses was (12.2+/-7.5)%. The MOSFET measured patient skin doses also had good reproducibility, with inter-fraction deviations ranging from 1.4% to 6.6%. Similar results were found from trials using a RANDO phantom.
CONCLUSIONS: The MOSFETs performed well when used in the tomotherapy Hi-Art unit and did not increase the overall treatment set-up time when used for patient measurements. It was found that MOSFETs are suitable detectors for surface dose measurements in both conventional beam and tomotherapy treatments and they can provide valuable skin dose information in areas where the treatment planning system may not be accurate.

Entities:  

Mesh:

Year:  2007        PMID: 18037519     DOI: 10.1016/j.radonc.2007.10.025

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  3 in total

1.  Acute Toxicity From Breast Cancer Radiation Using Helical Tomotherapy With a Simultaneous Integrated Boost.

Authors:  Andrzej P Wojcieszynski; Anna K Olson; Yi Rong; Randall J Kimple; Poonam Yadav
Journal:  Technol Cancer Res Treat       Date:  2015-03-16

2.  Peripheral dose measurement in high-energy photon radiotherapy with the implementation of MOSFET.

Authors:  Vassiliki Vlachopoulou; Georgia Malatara; Harry Delis; Kiki Theodorou; Dimitrios Kardamakis; George Panayiotakis
Journal:  World J Radiol       Date:  2010-11-28

3.  Evaluation of the usefulness of a MOSFET detector in an anthropomorphic phantom for 6-MV photon beam.

Authors:  Ryosuke Kohno; Eriko Hirano; Satoshi Kitou; Tomonori Goka; Kana Matsubara; Satoru Kameoka; Taeko Matsuura; Takaki Ariji; Teiji Nishio; Mitsuhiko Kawashima; Takashi Ogino
Journal:  Radiol Phys Technol       Date:  2010-02-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.