Attempt to improve transcutaneous bilirubinometry: a double-blind study of Medick BiliMed versus Respironics BiliCheck.

D De Luca1, E Zecca, M Corsello, E Tiberi, C Semeraro, C Romagnoli.   

Abstract

OBJECTIVES: To compare the accuracy of a new transcutaneous bilirubinometer, BiliMed (Medick SA, Paris, France) with BiliCheck (Respironics, Marietta, GA, USA), a widely available instrument, and with total serum bilirubin measurement.
DESIGN: A prospective double-blind study comparing the two devices was carried out. 686 healthy newborns needing measurement of their bilirubin were enrolled over a 4-month period. Serum and transcutaneous bilirubin measurements were taken with both devices within 15 minutes. The order of use of the instruments was randomised.
SETTING: Well-baby nursery ward in a university hospital, tertiary referral centre.
RESULTS: The linear regression analysis showed a better correlation between BiliCheck and serum bilirubin (r = 0.75) than between BiliMed and serum bilirubin (r = 0.45). BiliCheck variability (+/-2 SD of the mean bias from serum bilirubin) was within -87.2 to 63.3 micromol/l, while BiliMed variability was within -97.5 to 121.4 micromol/l. The receiver operating characteristic analysis (for serum bilirubin levels >205.2 micromol/l or >239.4 micromol/l) showed significantly higher areas under the curve for BiliCheck than those for BiliMed (p<0.001).
CONCLUSIONS: Despite the potential practical advantages of BiliMed, its reduced diagnostic accuracy in comparison with BiliCheck does not justify its use in clinical practice.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18006568     DOI: 10.1136/adc.2007.121053

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  5 in total

1.  Transcutaneous bilirubin measurement at the time of hospital discharge in a multiethnic newborn population.

Authors:  Douglas M Campbell; Karoon C Danayan; Valleverdina McGovern; Sohail Cheema; Brenda Stade; Michael Sgro
Journal:  Paediatr Child Health       Date:  2011-03       Impact factor: 2.253

2.  Stroke volume and cardiac output evaluation by electrical cardiometry: accuracy and reference nomograms in hemodynamically stable preterm neonates.

Authors:  A Boet; G Jourdain; S Demontoux; D De Luca
Journal:  J Perinatol       Date:  2016-04-21       Impact factor: 2.521

3.  Measuring transcutaneous bilirubin: a comparative analysis of three devices on a multiracial population.

Authors:  Francesco Raimondi; Silvia Lama; Francesca Landolfo; Maria Sellitto; Angela Carla Borrelli; Rosalba Maffucci; Paola Milite; Letizia Capasso
Journal:  BMC Pediatr       Date:  2012-06-14       Impact factor: 2.125

4.  Large scale validation of a new non-invasive and non-contact bilirubinometer in neonates with risk factors.

Authors:  Animesh Halder; Aniruddha Adhikari; Ria Ghosh; Soumendra Singh; Amrita Banerjee; Nilanjana Ghosh; Arnab Madhab Bhattacharya; Shrabani Mandal; Prantar Chakrabarti; Debasis Bhattacharyya; Hatem M Altass; Moataz Morad; Saleh A Ahmed; Asim Kumar Mallick; Samir Kumar Pal
Journal:  Sci Rep       Date:  2020-07-07       Impact factor: 4.379

5.  Comparison of a new transcutaneous bilirubinometer (Bilimed) with serum bilirubin measurements in preterm and full-term infants.

Authors:  Tanja Karen; Hans Ulrich Bucher; Jean-Claude Fauchère
Journal:  BMC Pediatr       Date:  2009-11-12       Impact factor: 2.125

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.