Literature DB >> 18005043

Influence of multiple clinical use on fatigue resistance of ProTaper rotary nickel-titanium instruments.

E P Vieira1, E C França, R C Martins, V T L Buono, M G A Bahia.   

Abstract

AIM: To examine the influence of clinical use on the occurrence of deformation and fracture and on the fatigue resistance of ProTaper rotary instruments.
METHODOLOGY: Root canal treatments were performed on patients using the ProTaper rotary system. Ten sets of instruments were used by an experienced endodontist, each set in five molars. Another 10 sets of instruments were used by the same operator, each set in eight molars. In addition, 10 sets of instruments were used, each set in five molars, by undergraduate students with no clinical experience with the system. After clinical use, S1, S2, F1 and F2 instruments were analysed for damage by optical and scanning electron microscopy. The used sets, along with a control group of 12 sets of new instruments, were then tested in a bench device for fatigue resistance.
RESULTS: The use of the ProTaper rotary instruments by an experienced endodontist allowed for the cleaning and shaping of the root canal system of up to eight molars without fracture. During the students work, six instruments fractured. Fatigue resistance decreased upon clinical use for all instruments analysed.
CONCLUSIONS: Fatigue resistance of used instruments was reduced, but no significant change was observed amongst the instruments used for shaping the canals of five and eight molars. Operator experience affected the occurrence of fracture and plastic deformation during shaping.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18005043     DOI: 10.1111/j.1365-2591.2007.01336.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  11 in total

1.  Influence of cyclic loading in NiTi austenitic and R-phase endodontic files from a finite element perspective.

Authors:  Suzanny C S Martins; Jessica D Silva; Paula R Garcia; Ana C D Viana; Vicente T L Buono; Leandro A Santos
Journal:  Clin Oral Investig       Date:  2022-01-17       Impact factor: 3.573

2.  Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel-titanium endodontic instruments.

Authors:  Felipe Augusto Restrepo-Restrepo; Viviana Andrea Holguín-Vásquez; Syldana Julieth Cañas-Jiménez; Paula Andrea Villa-Machado; Sara Ochoa-Soto; Claudia Patricia Ossa-Orozco; Sergio Iván Tobón-Arroyave
Journal:  Dent Res J (Isfahan)       Date:  2021-06-22

3.  Evaluation of deformation and fracture rates for nickel-titanium rotary instruments according to the frequency of clinical use.

Authors:  Ayça Yılmaz; Sıtkı Selçuk Gökyay; Rüştü Dağlaroğlu; Işıl Karagöz Küçükay
Journal:  Eur Oral Res       Date:  2018-05-01

4.  Cyclic fatigue resistance of Mtwo NiTi rotary instruments used by experienced and novice operators--an in vivo and in vitro study.

Authors:  Gianluca Plotino; Dina Al-Sudani; Stefano Pulino; Nicola M Grande; Piero Alessandro Marcoli; Silvia Pizzi; Luca Testarelli; Gianluca Gambarini
Journal:  Med Sci Monit       Date:  2012-06

5.  Importance of rotary systems in dental care by undergraduate students in patients of a public health service of Belo Horizonte.

Authors:  Fernanda-Alcaraz-Orta Bruno; Eduardo Nunes; Martinho-Campolina-Rebello Horta; Ana-Maria-Abras da Fonseca; Frank-Ferreira Silveira
Journal:  J Clin Exp Dent       Date:  2016-02-01

6.  Cyclic flexural fatigue resistance of NiTi Controlled Memory and Blue Technology instruments after torsional preloading.

Authors:  Túlio César Modesto; Eufemia Carolina Peláez Acosta; Pedro Damas Resende; Érika Sales Joviano Pereira; Isabella Faria da Cunha Peixoto; Vicente Tadeu Lopes Buono; Ana Cecília Diniz Viana
Journal:  J Appl Oral Sci       Date:  2018-08-20       Impact factor: 2.698

7.  A comparative study of root canal shaping using protaper universal and protaper next rotary files in preclinical dental education.

Authors:  Gül Çelik; Feyza Özdemir Kısacık; Emir Faruk Yılmaz; Arife Mersinlioğlu; İhsan Furkan Ertuğrul; Hikmet Orhan
Journal:  PeerJ       Date:  2019-08-19       Impact factor: 2.984

8.  Reciproc Endodontic File Surface Defects After Single Use: An SEM Analysis.

Authors:  Mohammed Howait
Journal:  J Int Soc Prev Community Dent       Date:  2021-01-30

9.  Failure of ProTaper rotary Ni-Ti instruments used by undergraduate students.

Authors:  Marc García-Font; Fernando Duran-Sindreu; Sergio Morello-Castro; Montserrat Mercade-Bellido; Rufino Bueno-Martínez; Miguel Roig-Cayón
Journal:  J Clin Exp Dent       Date:  2012-10-01

10.  Effects of R-Phase on Mechanical Responses of a Nickel-Titanium Endodontic Instrument: Structural Characterization and Finite Element Analysis.

Authors:  Leandro de Arruda Santos; Pedro Damas Resende; Maria Guiomar de Azevedo Bahia; Vicente Tadeu Lopes Buono
Journal:  ScientificWorldJournal       Date:  2016-05-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.