Russell P Dresher1, Pedro P Irazoqui.
Abstract
Wireless sensing has shown potential benefits for the continuous-time measurement of physiological data. One such application is the recording of intraocular pressure (IOP) for patients with glaucoma. Ultra-low-power circuits facilitate the use of inductively-coupled power for implantable wireless systems. Compact circuit size is also desirable for implantable systems. As a first step towards the realization of such circuits, we have designed a compact, ultra-low-power operational amplifier which can be used to record IOP. This paper presents the measured results of a CMOS operational amplifier that can be incorporated with a wireless IOP monitoring system or other low-power application. It has a power consumption of 736 nW, chip area of 0.023 mm2, and output impedance of 69 Omega to drive low-impedance loads.Entities:
Mesh:
Year: 2007 PMID: 18003395 DOI: 10.1109/IEMBS.2007.4353729