Literature DB >> 18002976

A 128-electrode three dimensional electrical impedance tomography system.

Guizhi Xu1, Renping Wang, Shuai Zhang, Shuo Yang, Gusphyl A Justin, Mingui Sun, Weili Yan.   

Abstract

Electrical impedance tomography (EIT) is a new functional imaging technique. This paper presents the development of a new electrical impedance tomography system with 128 electrodes for impedance change detection and 3D imaging of the human thorax. The system consists of several modules, including multi-frequency current source, driving, measuring, data acquisition, and controlling and signal processing modules. A high speed digital signal processor (DSP) is used as the controller. The 64 driving electrodes and 64 measuring electrodes are positioned uniformly in four planes around the surface of a cylindrical phantom filled with a saline solution and objects of varying conductivities. The performance has been tested, and these preliminary experiments demonstrate the feasibility this system.

Entities:  

Mesh:

Year:  2007        PMID: 18002976     DOI: 10.1109/IEMBS.2007.4353310

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Electrical impedance tomography measured at two thoracic levels can visualize the ventilation distribution changes at the bedside during a decremental positive end-expiratory lung pressure trial.

Authors:  Ido G Bikker; Carsten Preis; Mahamud Egal; Jan Bakker; Diederik Gommers
Journal:  Crit Care       Date:  2011-08-11       Impact factor: 9.097

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.