Literature DB >> 18002834

A preliminary study to demonstrate the use of an air bag device to prevent fall-related injuries.

Toshiyo Tamura1, Takumi Yoshimura, Masaki Sekine.   

Abstract

We sought to develop a fall-related injury prevention system using a telemetry acceleration monitor and an air bag. In a preliminary study, we developed a fall detection system using an acceleration signal and triggering the inflation of an air bag. Sixteen subjects mimicked falls and their acceleration waveforms were monitored. Then, we developed a fall detection algorithm that could detect signals 300 ms before the fall. This signal was used as a trigger to inflate the air bag to a capacity of 2.4 L. While the proposed system can aid in preventing fall-related injuries, further development is needed to miniaturize the inflation system.

Entities:  

Mesh:

Year:  2007        PMID: 18002834     DOI: 10.1109/IEMBS.2007.4353168

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X