Literature DB >> 17999668

Accuracy of pulse oximetry in assessing heart rate of infants in the neonatal intensive care unit.

Jasbir K S B Singh1, C Omar F Kamlin, Colin J Morley, Colm P F O'Donnell, Susan M Donath, Peter G Davis.   

Abstract

AIM: To determine the accuracy of pulse oximetry measurement of heart rate in the neonatal intensive care unit.
METHODS: Stable preterm infants were monitored with a pulse oximeter and an ECG. The displays of both monitors were captured on video. Heart rate data from both monitors, including measures of signal quality, were extracted and analysed using Bland Altman plots.
RESULTS: In 30 infants the mean (SD) difference between heart rate measured by pulse oximetry and electrocardiography was -0.4 (12) beats per minute. Accuracy was maintained when the signal quality or perfusion was low.
CONCLUSION: Pulse oximetry may provide a useful measurement of heart rate in the neonatal intensive care unit. Studies of this technique in the delivery room are indicated.

Entities:  

Mesh:

Year:  2007        PMID: 17999668     DOI: 10.1111/j.1440-1754.2007.01250.x

Source DB:  PubMed          Journal:  J Paediatr Child Health        ISSN: 1034-4810            Impact factor:   1.954


  6 in total

1.  Smartphone app for neonatal heart rate assessment: an observational study.

Authors:  Susanna Myrnerts Höök; Nicolas J Pejovic; Francesco Cavallin; Clare Lubulwa; Josaphat Byamugisha; Jolly Nankunda; Thorkild Tylleskär; Tobias Alfven
Journal:  BMJ Paediatr Open       Date:  2020-07-22

2.  Avoiding hyperoxemia during neonatal resuscitation: time to response of different SpO2 monitors.

Authors:  Hernando Baquero; Ramiro Alviz; Armando Castillo; Fredy Neira; Augusto Sola
Journal:  Acta Paediatr       Date:  2011-01-17       Impact factor: 2.299

3.  A conversion formula for comparing pulse oximeter desaturation rates obtained with different averaging times.

Authors:  Jan Vagedes; Anja Bialkowski; Cornelia Wiechers; Christian F Poets; Klaus Dietz
Journal:  PLoS One       Date:  2014-01-28       Impact factor: 3.240

4.  Electrocardiography versus photoplethysmography in assessment of maternal heart rate variability during labor.

Authors:  Hernâni Gonçalves; Paula Pinto; Manuela Silva; Diogo Ayres-de-Campos; João Bernardes
Journal:  Springerplus       Date:  2016-07-15

5.  The perfusion index of healthy term infants during transition at birth.

Authors:  Jacco K Kroese; Jeroen J van Vonderen; Ilona C Narayen; Frans J Walther; Stuart Hooper; Arjan B te Pas
Journal:  Eur J Pediatr       Date:  2015-10-24       Impact factor: 3.183

6.  A Textile Sleeve for Monitoring Oxygen Saturation Using Multichannel Optical Fibre Photoplethysmography.

Authors:  Hattan K Ballaji; Ricardo Correia; Serhiy Korposh; Barrie R Hayes-Gill; Francisco U Hernandez; Byron Salisbury; Stephen P Morgan
Journal:  Sensors (Basel)       Date:  2020-11-17       Impact factor: 3.576

  6 in total