Dosimetric and qualitative analysis of kinetic properties of millennium 80 multileaf collimator system for dynamic intensity modulated radiotherapy treatments.

Anup K Bhardwaj1, T S Kehwar, S K Chakarvarti, A S Oinam, S C Sharma.   

Abstract

The aim of this paper is to analyze the positional accuracy, kinetic properties of the dynamic multileaf collimator (MLC) and dosimetric evaluation of fractional dose delivery for the intensity modulated radiotherapy (IMRT) for step and shoot and sliding window (dynamic) techniques of Varian multileaf collimator millennium 80. Various quality assurance tests such as accuracy in leaf positioning and speed, stability of dynamic MLC output, inter and intra leaf transmission, dosimetric leaf separation and multiple carriage field verification were performed. Evaluation of standard field patterns as pyramid, peaks, wedge, chair, garden fence test, picket fence test and sweeping gap output was done. Patient dose quality assurance procedure consists of an absolute dose measurement for all fields at 5 cm depth on solid water phantom using 0.6 cc water proof ion chamber and relative dose verification using Kodak EDR-2 films for all treatment fields along transverse and coronal direction using IMRT phantom. The relative dose verification was performed using Omni Pro IMRT film verification software. The tests performed showed acceptable results for commissioning the millennium 80 MLC and Clinac DHX for dynamic and step and shoot IMRT treatments.

Entities:  

Mesh:

Year:  2007        PMID: 17998715     DOI: 10.4103/0973-1482.31967

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  3 in total

1.  Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D.

Authors:  S A Rohani; S R Mahdavi; A Mostaar; S Ueltzhöffer; R Mohammadi; Gh Geraily
Journal:  J Biomed Phys Eng       Date:  2019-02-01

2.  An Investigation of Multileaf Collimator Performance Dependence on Gantry Angle Using Machine Log Files.

Authors:  Jeremy L Hughes; Martin A Ebert; Conor K McGarry; Christina E Agnew; Mahsheed Sabet; Pejman Rowshanfarzad
Journal:  J Med Phys       Date:  2021-12-31

3.  Quality assurance of MLC leaf position accuracy and relative dose effect at the MLC abutment region using an electronic portal imaging device.

Authors:  Iori Sumida; Hajime Yamaguchi; Hisao Kizaki; Masahiko Koizumi; Toshiyuki Ogata; Yutaka Takahashi; Yasuo Yoshioka
Journal:  J Radiat Res       Date:  2012-07-10       Impact factor: 2.724

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.