PURPOSE: To present a novel technique to deliver the endothelial graft in small incision Descemet Stripping Automated Endothelial Keratoplasty (DSAEK) and to compare graft trauma between forceps delivery and the novel technique.
DESIGN: Laboratory investigation with an interventional case report.
METHODS: Ten pairs of human donor corneas were sectioned using an automated microkeratome system (Moria ALTK System, Antony, France). The recipient model was prepared by creating a 3 mm clear corneal wound in another human donor cornea. For each pair of corneas, one endothelial graft underwent single-fold delivery with forceps, while the other was delivered with the novel cartridge based technique. Each graft was stained with 0.25% trypan blue and 0.2% alizarin red, and digital photomicrographs were taken. A proportion of graft injury was calculated and differences were analyzed. Subsequently, a patient requiring DSAEK underwent the modified novel insertion technique.
RESULTS: After insertion, the mean proportion of graft endothelial injury from forceps delivery through the ex vivo model was 26.02% (n = 10, standard deviation [SD] +/- 14.85%). The mean proportion of graft endothelial injury from cartridge delivery was 9.85% (n = 10, SD +/- 4.33%). The median difference between the two methods was -13%, representing less endothelial injury with the cartridge. This difference was statistically significant (P = .006). The patient who underwent DSAEK with this technique had improved visual acuity and a clear graft at five months.
CONCLUSIONS: In our surgical model, inserting an endothelial graft through a small corneal wound using a novel cartridge-based technique created significantly less endothelial damage than with forceps insertion. Clinically, this technique was performed without complication.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17996209     DOI: 10.1016/j.ajo.2007.08.036

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  8 in total

Review 1.  Graft survival and endothelial outcomes in the new era of endothelial keratoplasty.

Authors:  Sanjay V Patel
Journal:  Exp Eye Res       Date:  2011-06-15       Impact factor: 3.467

Review 2.  [Critical endothelial procedures during posterior lamellar graft preparation and transplantation].

Authors:  T A Fuchsluger; F E Kruse; G Geerling
Journal:  Ophthalmologe       Date:  2017-08       Impact factor: 1.059

3.  Effects of a novel push-through technique using the implantable collamer lens injector system for graft delivery during endothelial keratoplasty.

Authors:  Sug Jae Kang; Myung Hun Kim; Mee Kum Kim; Won Ryang Wee; Jin Hak Lee; Eui Sang Chung
Journal:  Korean J Ophthalmol       Date:  2013-02-28

4.  Clinicopathologic findings in failed descemet stripping automated endothelial keratoplasty.

Authors:  Qing Zhang; J Bradley Randleman; R Doyle Stulting; W Barry Lee; Donald U Stone; Alan M Kozarsky; Hans E Grossniklaus
Journal:  Arch Ophthalmol       Date:  2010-08

5.  Anomalous systemic arterial supply to the left basal lung with a calcified abnormal vessel: a case report.

Authors:  Motoki Ebihara; Sakashi Fujimori; Souichiro Suzuki; Takuma Yotsumoto; Shinichiro Kikunaga; Reo Ohtsuka; Shigefumi Matsuyama
Journal:  Surg Case Rep       Date:  2022-06-22

6.  Interleukin-1β enhances cell migration through AP-1 and NF-κB pathway-dependent FGF2 expression in human corneal endothelial cells.

Authors:  Jeong Goo Lee; Martin Heur
Journal:  Biol Cell       Date:  2013-03-07       Impact factor: 4.458

7.  Real-time assessment of corneal endothelial cell damage following graft preparation and donor insertion for DMEK.

Authors:  Maninder Bhogal; Chan N Lwin; Xin-Yi Seah; Elavazhagan Murugan; Khadijah Adnan; Shu-Jun Lin; Gary Peh; Jodhbir S Mehta
Journal:  PLoS One       Date:  2017-10-04       Impact factor: 3.240

8.  Suture pull-through insertion of graft donor in Descemet stripping automated endothelial keratoplasty: Results of 4-year follow-up.

Authors:  Vincenzo Sarnicola; Chiara Millacci; Enrica Sarnicola; Caterina Sarnicola; Francesco Sabatino; Andrew Ruggiero
Journal:  Taiwan J Ophthalmol       Date:  2015-08-28
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.