Exposure to noise during continuous positive airway pressure: influence of interfaces and delivery systems.

F Cavaliere1, G Conti, R Costa, G Spinazzola, R Proietti, A Sciuto, S Masieri.   

Abstract

BACKGROUND: We measured noise intensity and perceived noisiness during continuous positive airway pressure (CPAP) performed with two interfaces (face-mask, helmet) and four delivery systems.
METHODS: Eight healthy volunteers received CPAP in random order with: two systems provided with a flow generator using the Venturi effect and a mechanical expiratory valve (A: Venturi, Starmed; B: Whisperflow-2, Caradyne Ltd); one 'free-flow' system provided with high flow O(2) and air flowmeters, an inspiratory gas reservoir, and a water valve (C: CF800, Drägerwerk, AG); and a standard mechanical ventilator (Servoventilator 300, Siemens-Elema). Systems A, B, and C were tested with a face-mask and a helmet at a CPAP value of 10 cm H(2)O; the mechanical ventilator was only tested with the face mask. Noise intensity was measured with a sound-level meter. After each test, participants scored noisiness on a visual analog scale (VAS).
RESULTS: The noise levels measured ranged from 57+/-11 dBA (mechanical ventilator plus mask) to 93+/-1 and 94+/-2 dBA (systems A and B plus helmet) and were significantly affected by CPAP systems (A and B noisier than C and D) and interfaces (helmet CPAP noisier than mask CPAP). Subjective evaluation showed that systems A and B plus helmet were perceived as noisier than system C plus mask or helmet.
CONCLUSIONS: Maximum noise levels observed in this study may potentially cause patient discomfort. Less noisy CPAP systems (not using Venturi effect) and interfaces (facial mask better than helmet) should be preferred, particularly for long or nocturnal treatments.

Entities:  

Mesh:

Year:  2007        PMID: 17996003     DOI: 10.1111/j.1399-6576.2007.01474.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  8 in total

1.  Should we carry out noninvasive ventilation using a helmet in acute respiratory distress syndrome?

Authors:  Rémi Coudroy; Jean-Pierre Frat; Arnaud W Thille
Journal:  Ann Transl Med       Date:  2016-09

2.  Acute Respiratory Failure in COVID-19 Pneumoniae: The Best Interface for an Optimal Management - A South Italian Experience.

Authors:  Pasquale Imitazione; Giorgio Emanuele Polistina; Anna Annunziata; Valentina Di Spirito; Maurizia Lanza; Giuseppe Fiorentino
Journal:  Turk Thorac J       Date:  2020-06-10

3.  Impact of a New Nasal Pillows Mask on Patients' Acceptance, Compliance, and Willingness to Remain on CPAP Therapy.

Authors:  Alison Wimms; Sahisha Ketheeswaran; Claus Ziegenbein; Laura Jennings; Holger Woehrle
Journal:  Sleep Disord       Date:  2016-08-25

4.  Effect of Filters on the Noise Generated by Continuous Positive Airway Pressure Delivered via a Helmet.

Authors:  Ricardo Hernández-Molina; Francisco Fernández-Zacarías; Isabel Benavente-Fernández; Gema Jiménez-Gómez; Simón Lubián-López
Journal:  Noise Health       Date:  2017 Jan-Feb       Impact factor: 0.867

5.  Investigating Noise Exposure to Newborn Infants From Respiratory Support: Methodological Considerations.

Authors:  Devika Singh; Gerhard Fusch
Journal:  Cureus       Date:  2021-11-08

Review 6.  Helmet Ventilation for Pediatric Patients During the COVID-19 Pandemic: A Narrative Review.

Authors:  Shu-Chi Mu; Yu-Hsuan Chien; Pin-Zhen Lai; Ke-Yun Chao
Journal:  Front Pediatr       Date:  2022-02-02       Impact factor: 3.418

Review 7.  Role of helmet ventilation during the 2019 coronavirus disease pandemic.

Authors:  Ke-Yun Chao; Jong-Shyan Wang; Wei-Lun Liu
Journal:  Sci Prog       Date:  2022 Apr-Jun       Impact factor: 2.774

8.  High flow biphasic positive airway pressure by helmet--effects on pressurization, tidal volume, carbon dioxide accumulation and noise exposure.

Authors:  Onnen Moerer; Peter Herrmann; José Hinz; Paolo Severgnini; Edoardo Calderini; Michael Quintel; Paolo Pelosi
Journal:  Crit Care       Date:  2009-06-05       Impact factor: 9.097

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.