Literature DB >> 17994881

Three-dimensional measurement endoscope system with virtual rulers.

Hiromasa Nakatani1, Keiichi Abe, Atsuo Miyakawa, Susumu Terakawa.   

Abstract

Conventional endoscopic images do not provide quantitative 3-D information. We present an endoscope system that can measure the size and position of an object in real time. Our endoscope contains four laser beam sources and a camera. The procedural steps for 3-D measurements are as follows. First, to obtain the function that maps 2-D coordinates of an image point to its 3-D coordinates in 3-D space, we observe a standard chart with the endoscope lens and determine the correspondence between the image and object height. In addition to the mapping, this function can correct barrel-shaped distortion of endoscopic images. The system detects laser spots on an object surface automatically using a template matching method, and maps the 2-D coordinates of the laser spots to the 3-D coordinates by the triangulation method. Then the system calculates the magnification ratio on the object plane, which is perpendicular to the optical axis and passes the laser spot, so that the system can superimpose a ruler whose scale fits the 3-D coordinates of the object. Thus, physicians can measure the size and position of objects in real time on undistorted images similar to placing rulers on the surface of an organ.

Entities:  

Mesh:

Year:  2007        PMID: 17994881     DOI: 10.1117/1.2800758

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  3 in total

1.  Feasibility of obtaining quantitative 3-dimensional information using conventional endoscope: a pilot study.

Authors:  Jong Jin Hyun; Hoon Jai Chun; Bora Keum; Yeon Seok Seo; Yong Sik Kim; Yoon Tae Jeen; Hong Sik Lee; Soon Ho Um; Chang Duck Kim; Ho Sang Ryu; Jong-Wook Lim; Dong-Gi Woo; Young-Joong Kim; Myo-Taeg Lim
Journal:  Clin Endosc       Date:  2012-08-22

2.  Recovering 3D Shape with Absolute Size from Endoscope Images Using RBF Neural Network.

Authors:  Seiya Tsuda; Yuji Iwahori; M K Bhuyan; Robert J Woodham; Kunio Kasugai
Journal:  Int J Biomed Imaging       Date:  2015-04-09

3.  Integrated Multipoint-Laser Endoscopic Airway Measurements by Transoral Approach.

Authors:  Marie Neitsch; Iris-Susanne Horn; Mathias Hofer; Andreas Dietz; Miloš Fischer
Journal:  Biomed Res Int       Date:  2016-02-15       Impact factor: 3.411