Elizabeth Greiner-Sosanko1, Spiros Giannoutsos, Darla R Lower, Mohamed A Virji, Matthew D Krasowski.
Abstract
A high-performance liquid chromatography (HPLC) assay using UV detection is described for the simultaneous measurement of the newer generation anti-epileptic medications lamotrigine, oxcarbazepine (parent drug and active metabolite 10- hydroxycarbazepine), and zonisamide. Detection of all four compounds can be done at 230 nm; however, there is a potential interference with zonisamide in patients on clonazepam therapy. Therefore, the method uses dual wavelength detection: 230 nm for oxcarbazepine and 10-hydroxycarbazepine and 270 nm for lamotrigine and zonisamide. In addition, a simple gas chromatography method using a nitrogen-phosphorus detector is described for the measurement of levetiracetam, another of the recently approved anti-epileptic medications. For both methods, limits of quantitation, linearities, accuracies, and imprecisions cover the therapeutic range for drug monitoring of patients. A wide variety of clinical drugs, including other anti-epileptic drugs, do not interfere with these assays. These procedures would be of special interest to clinical laboratories, particularly due to the limited availability of immunoassays for newer generation anti-epileptic medications and that therapeutic uses of these drugs are expanding beyond epilepsy to other neurologic and psychiatric disorders.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17988451 PMCID: PMC2231334 DOI: 10.1093/chromsci/45.9.616
Source DB: PubMed Journal: J Chromatogr Sci ISSN: 0021-9665 Impact factor: 1.618