A biomechanical analysis of a novel arthroscopic suture method compared to standard suture knots and materials for rotator cuff repair.

Andrew Mahar1, Tim Odell, Walter Thomas, Robert Pedowitz.   

Abstract

PURPOSE: To biomechanically compare a novel suture welding method to 2 standard suture materials when tied with 3 types of arthroscopic knots.
METHODS: Sixty suture loops were tied with Weston, Roeder, and Duncan knots for both No. 2 FiberWire (FW) and No. 2 Ethibond (EB; 10 loops/group). Ten suture loops were created using a novel technique using No. 2 Nylon suture, with welding of the suture ends using sonic energy. All loops were created over the same dowel. Loops were cycled between 10N and 45N for 200 cycles, followed by a failure test. Loop elongation and ultimate failure load were compared between welded sutures and both materials using a 2-way analysis of variance.
RESULTS: There were no differences in loop elongation for the welded Nylon sutures compared to EB Roeder and EB Duncan knots, although EB Weston knots had significantly less loop elongation (P < .05). Loop elongation was significantly lower with FW than with EB (P < .001). For maximum failure load, there were no differences between the various EB loops. Welded loops demonstrated similar maximum failure loads to EB Roeder and EB Duncan knots, but were lower than EB Weston knots. Overall, the FW knots were significantly stronger than the EB groups (P < .001 for all comparisons).
CONCLUSIONS: A novel method to simplify the knot tying procedure by welding the suture ends demonstrated mechanical properties generally similar to No. 2 Ethibond. This method provides: (1) surgical advantages by eliminating knot tying; (2) a quick and easy locking mechanism; and (3) a low profile that eliminates high-profile knots. However, it is not clear whether the lower stiffness or the lower failure load may detract from these potential benefits compared to constructs with greater stiffness. CLINICAL RELEVANCE: The suture welding technique generally behaved in a manner similar to that of Ethibond suture loops, although the difficulty of tying arthroscopic knots was eliminated.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17986402     DOI: 10.1016/j.arthro.2007.06.011

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  3 in total

1.  Loop security and tensile properties of polyblend and traditional suture materials.

Authors:  Mehmet Türker; Onder Kılıçoğlu; Ahmet Salduz; Ergun Bozdağ; Emin Sünbüloğlu
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-06-10       Impact factor: 4.342

2.  Is suture comparable to wire for cerclage fixation? A biomechanical analysis.

Authors:  Scott E Westberg; Yves P Acklin; Siva Hoxha; Cagri Ayranci; Samer Adeeb; Martin Bouliane
Journal:  Shoulder Elbow       Date:  2017-10-12

3.  Biomechanical evaluation of the Nice knot.

Authors:  Shannon W Hill; Christopher R Chapman; Samer Adeeb; Kajsa Duke; Lauren Beaupre; Martin J Bouliane
Journal:  Int J Shoulder Surg       Date:  2016 Jan-Mar
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.