| Literature DB >> 17979441 |
D K Wood1, M V Requa, A N Cleland.
Abstract
We describe the design, fabrication, and use of a radio frequency reflectometer integrated with a microfluidic system, applied to the very high-throughput measurement of micron-scale particles, passing in a microfluidic channel through the sensor region. The device operates as a microfabricated Coulter counter [U.S. Patent No. 2656508 (1953)], similar to a design we have described previously, but here with significantly improved electrode geometry as well as including electronic tuning of the reflectometer; the two improvements yielding an improvement by more than a factor of 10 in the signal to noise and in the diametric discrimination of single particles. We demonstrate the high-throughput discrimination of polystyrene beads with diameters in the 4-10 microm range, achieving diametric resolutions comparable to the intrinsic spread of diameters in the bead distribution, at rates in excess of 15 x 10(6) beads/h.Entities:
Mesh:
Year: 2007 PMID: 17979441 DOI: 10.1063/1.2794230
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523