Literature DB >> 17976109

Inappropriate defibrillator shocks from depolarization--repolarization mismatch in a patient with hypertrophy cardiomyopathy.

Suraj Kapa1, Jay H Curwin, Robert F Coyne, Stephen L Winters.   

Abstract

Despite wide use of dedicated bipolar sensing electrodes in implantable cardioverter-defibrillator (ICD) systems, oversensing occasionally occurs, leading to unwarranted shocks or antitachycardia pacing. This case report highlights an individual with hypertrophic cardiomyopathy (HCM) who experienced inappropriate shocks from oversensing of repolarization electrograms (T-waves). During the implantation procedure, no excessive T-wave amplitudes were detected during sinus rhythm, ventricular pacing, or induced ventricular fibrillation. T-wave oversensing leading to shocks only developed after maturation of the lead-tissue interface. An adequate safety margin for discrimination between ventricular electrograms and T-waves could not be assured. Thus, insertion of a new dedicated pacing-sensing electrode was required. The degree to which intracardiac repolarization signals may be heightened in patients with HCM has not been investigated systematically. However, a relative decrease in the ventricular electrogram amplitude without a concomitant decline of the intracardiac T-wave amplitude appears to have led to the problem in this patient. Special caution in technique and device selection with a particular emphasis on T-wave sensing may be prudent when ICDs are implanted in individuals with HCM. Additional programmable variables may also be beneficial in such cases.

Entities:  

Mesh:

Year:  2007        PMID: 17976109     DOI: 10.1111/j.1540-8159.2007.00881.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Treatment of symptomatic bradycardia due to T-wave oversensing with implantation of a new generator incorporating delayed decay and threshold start sensitization algorithms.

Authors:  Karim Nathan; Omnia Kamel; Daniel Marks; Andrea M Russo
Journal:  HeartRhythm Case Rep       Date:  2020-07-01
  1 in total