Literature DB >> 17974120

A simplified method to assess precision of fit between framework and supporting implants: a preliminary study.

Angelo Calderini1, Carlo Maiorana, Giuliano Garlini, Tommaso Abbondanza.   

Abstract

PURPOSE: To present a simplified method for assessment of precision of fit between screw-retained frameworks and supporting implants in clinical situations.
MATERIALS AND METHODS: Torque-angle signature analysis is a method for analysis of tension within a joint. The OsseoCare device developed by Nobel Biocare was utilized for the tests. Three titanium frameworks were fabricated on the same master cast according to 3 different methods: (1) 1-piece casting, (2) the CNC (computer numeric controlled) method, and (3) the CTiP (Cresco Ti Precision) method. These frameworks were used to test the device and indirectly the application of the torque-angle signature analysis technology.
RESULTS: The frameworks fabricated according to the CNC and CTiP methods demonstrated OsseoCare tracings similar to the reference models of passively fitting joints, while the 1-piece cast framework did not.
CONCLUSION: According to this pilot study, the OsseoCare device and torque-angle signature analysis proved to be feasible for clinical assessment of fit between frameworks and supporting implants.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17974120

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  2 in total

1.  Vertical misfit of laser-sintered and vacuum-cast implant-supported crown copings luted with definitive and temporary luting agents.

Authors:  Raquel Castillo-de-Oyagüe; Andrés Sánchez-Turrión; José-Francisco López-Lozano; Alberto Albaladejo; Daniel Torres-Lagares; Javier Montero; Maria-Jesús Suárez-García
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2012-07-01

2.  A technique for in vitro fit assessment of multi-unit screw-retained implant restorations: Application of a triple-scan protocol.

Authors:  Stefan Holst; Matthias Karl; Manfred Wichmann; Ragai E Matta
Journal:  J Dent Biomech       Date:  2012-07-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.