Transcutaneous bilirubin in the pre-term infants.

Suwimol Sanpavat1, Issarang Nuchprayoon.   

Abstract

OBJECTIVE: 1) To evaluate the accuracy of transcutaneous bilirubin (T(C)B) measurement compared to total serum bilirubin level (TSB) in the pre-term infants and 2) To establish cut-off values of T(C)B that indicated need for TSB. MATERIAL AND
METHOD: Premature infants whose birth weight was more than 1,000 grams and gestational age less than 36 weeks had paired T(C)B-TSB assessment when jaundice was observed. T(C)B was done using JM 103 (Minolta AirShields Jaundice Meter) on the forehead. T(C)B, which corresponded to TSB level that required phototherapy, was chosen as the cut-off point that indicated blood sampling for TSB.
RESULTS: Two hundred and forty-nine paired T(CB)-TSB measurements from 196 premature neonates were obtained. Birth weight was 1,887 +/- 344.4 grams. TSB ranged from 4.5-17.6 mg/dL (mean 9.4, SD 2.2 mg/dL), T(C)B 4.1-17.7 mg/dL (mean 9.7, SD 2.4 mg/dL). The correlation coefficient between T(C)B and TSB was significant (r 0.79, p < 0.0001). T(C)B had a tendency to overestimate TSB with the mean difference of TSB- T(C)B = -0.3 +/- 1.5 mg/dL and 95% confidence interval of the mean -0.1 to -0.5 mg/dL. Of all the variables of birth weight, gestational age, and postnatal age, only postnatal age significantly influenced the correlation of T(C)B-TSB. In the early postnatal age of 1-4 days, the number of T(C)B reading overestimated TSB more than underestimated. However, at > or = 5 days of age, the number of underestimation was more than those of overestimation. The cut-off points of T(C)B that indicated the need for blood sampling for TSB were chosen as the same level of TSB of 6, 8, 10, 11, and 12 mg/dL when phototherapy was recommended. Screening with T(C)B would eliminate painful procedure of blood taking by 40%.
CONCLUSION: Noninvasive T(C)B assessment demonstrated significant accuracy when compared to TSB. T(C)B can be adopted as a screening test to identify the need for blood sampling of serum bilirubin in premature infants.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17957922

Source DB:  PubMed          Journal:  J Med Assoc Thai        ISSN: 0125-2208


  2 in total

1.  Screening neonatal jaundice based on the sclera color of the eye using digital photography.

Authors:  Terence S Leung; Karan Kapur; Ashley Guilliam; Jade Okell; Bee Lim; Lindsay W MacDonald; Judith Meek
Journal:  Biomed Opt Express       Date:  2015-10-23       Impact factor: 3.732

2.  Transcutaneous bilirubin level to predict hyperbilirubinemia in preterm neonates.

Authors:  Dewi Rahmawati; Mahendra Tri Arif Sampurna; Risa Etika; Martono Tri Utomo; Arend F Bos
Journal:  F1000Res       Date:  2020-04-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.