Literature DB >> 17951028

Influence of anisotropy on peri-implant stress and strain in complete mandible model from CT.

Sheng-Hui Liao1, Ruo-Feng Tong, Jin-Xiang Dong.   

Abstract

This paper reveals the influence of elastic anisotropy for the peri-implant stress and strain in personalized mandible. First, from CT data, the individual geometry of the complete range of mandible was well reproduced, also the separation between cortical and cancellous bone. Then, by an ad hoc automatic mesh generator integrated with anisotropic material assignment function, high quality anisotropic finite element model of the complete mandible was created, with two standard threaded implants embedded in posterior zone. The values of principal stress and strain in surrounding bone were evaluated under buccolingual oblique loading, and compared to that of the same FE model with equivalent isotropic material. Results of the analyses demonstrated that the percentage increase of stress and strain in anisotropic case reached up to 70%. It is concluded that anisotropy has significant effects on peri-implant stress and strain and careful consideration should be given to its use in biomechanical FE studies.

Entities:  

Mesh:

Year:  2007        PMID: 17951028     DOI: 10.1016/j.compmedimag.2007.09.001

Source DB:  PubMed          Journal:  Comput Med Imaging Graph        ISSN: 0895-6111            Impact factor:   4.790


  6 in total

1.  A high-fidelity 3D S-FEM stress analysis of a highly heterogeneous swine skull.

Authors:  S H Huo; C Jiang; X Cui; G R Liu
Journal:  Med Biol Eng Comput       Date:  2020-01-14       Impact factor: 2.602

2.  Comparison of ultimate force revealed by compression tests on extracted first premolars and FEA with a true scale 3D multi-component tooth model based on a CBCT dataset.

Authors:  Nuttapol Limjeerajarus; Phetcharat Dhammayannarangsi; Anon Phanijjiva; Pavita Tangsripongkul; Thanomsuk Jearanaiphaisarn; Pisha Pittayapat; Chalida Nakalekha Limjeerajarus
Journal:  Clin Oral Investig       Date:  2019-05-11       Impact factor: 3.573

3.  Influence of Trabecular Bone on Peri-Implant Stress and Strain Based on Micro-CT Finite Element Modeling of Beagle Dog.

Authors:  Sheng-Hui Liao; Xing-Hao Zhu; Jing Xie; Vikesh Kumar Sohodeb; Xi Ding
Journal:  Biomed Res Int       Date:  2016-06-14       Impact factor: 3.411

4.  Biomechanical Effects of Various Bone-Implant Interfaces on the Stability of Orthodontic Miniscrews: A Finite Element Study.

Authors:  Fabing Tan; Chao Wang; Chongshi Yang; Yuanding Huang; Yubo Fan
Journal:  J Healthc Eng       Date:  2017-06-19       Impact factor: 2.682

5.  The effect of mandibular flexure on the design of implant-supported fixed restorations of different facial types under two loading conditions by three-dimensional finite element analysis.

Authors:  Jing Gao; Xuejing Li; Jing He; Lulu Jiang; Baohong Zhao
Journal:  Front Bioeng Biotechnol       Date:  2022-08-29

6.  Micro-computed tomography derived anisotropy detects tumor provoked deviations in bone in an orthotopic osteosarcoma murine model.

Authors:  Heather A Cole; Tetsuro Ohba; Jiro Ichikawa; Jeffry S Nyman; Justin M M Cates; Hirotaka Haro; Herbert S Schwartz; Jonathan G Schoenecker
Journal:  PLoS One       Date:  2014-06-03       Impact factor: 3.240

  6 in total