Literature DB >> 17946001

An optical fiber photoplethysmographic system for central nervous system tissue.

Justin P Phillips1, Panayiotis A Kyriacou, Kuriakose J George, John V Priestley, Richard M Langford.   

Abstract

A new system for measuring the oxygen saturation of blood within tissue has been developed, for a number of potential patient monitoring applications. This proof of concept project aims to address the unmet need of real-time measurement of oxygen saturation in the central nervous system (CNS) for patients recovering from neurosurgery or trauma, by developing a fiber optic signal acquisition system for internal placement through small apertures. The development and testing of a two-wavelength optical fiber reflectance photoplethysmography (PPG) system is described. It was found that good quality red and near-infrared PPG signals could be consistently obtained from the human fingertip (n=6) and rat spinal cord (n=6) using the fiber optic probe. These findings justify further development and clinical evaluation of this fiber optic system.

Entities:  

Mesh:

Year:  2006        PMID: 17946001     DOI: 10.1109/IEMBS.2006.259690

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Cerebral arterial oxygen saturation measurements using a fiber-optic pulse oximeter.

Authors:  J P Phillips; R M Langford; S H Chang; K Maney; P A Kyriacou; D P Jones
Journal:  Neurocrit Care       Date:  2010-10       Impact factor: 3.210

  1 in total