| Literature DB >> 17945901 |
Roland Kempf1, Yuichi Kurita, Yoshichika Iida, Makoto Kaneko, Hiromu K Mishima, Hidetoshi Tsukamoto, Eiichiro Sugimoto.
Abstract
Non-contact tonometers are widely used to measure the internal eye pressure, i.e. the IntraOcular Pressure (IOP), which is an important parameter for the diagnosis and treatment of glaucoma. During the measurement, the eye is deformed by a short air pulse. Commonly the pressure dependent deformation is estimated from the time when the eye becomes flat, which is derived from the monitored reflection of an incident infrared light. We used a high speed camera to capture the complete motion of the eye directly and obtain more data during the pressure measurement. Assuming a simple eye model with non-linear material properties of the cornea, we extend our previous analysis of the motion of the eye, and obtain a similar principle shape of the eye deformation as observed in the experiments.Entities:
Mesh:
Year: 2006 PMID: 17945901 DOI: 10.1109/IEMBS.2006.259905
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X