Literature DB >> 17945880

Development of the re-entrant type resonant cavity applicator for brain tumor hyperthermia - experimental heating results.

T Yabuhara1, K Kato, K Tsuchiya, T Shigihara, T Uzuka, H Takahashi.   

Abstract

A re-entrant type resonant cavity applicator for brain tumor hyperthermia treatment is presented. In this method, a human head is placed between the gap of the inner re-entrant cylinders, and is heated with electromagnetic fields stimulated in the cavity without contact between the surface of the human head and the applicator. First, the comparison of the computer simulation and experimental results were presented. The estimated temperature agrees with the measured temperature with an error of 10% or better. Second, a method to control the heating pattern was presented. In the method, a human head was moved towards the electrode. The controllable range of heating pattern was about 70% of the distance from the center of the agar phantom in the direction of depth.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17945880     DOI: 10.1109/IEMBS.2006.259664

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Two 27 MHz Simple Inductive Loops, as Hyperthermia Treatment Applicators: Theoretical Analysis and Development.

Authors:  Vassilis Kouloulias; Irene Karanasiou; Maria Koutsoupidou; George Matsopoulos; John Kouvaris; Nikolaos Uzunoglu
Journal:  Comput Math Methods Med       Date:  2015-11-16       Impact factor: 2.238