Comparison between visual analysis and microscope assessment of surgical instrument cleanliness from sterile service departments.

I P Lipscomb1, A K Sihota, C W Keevil.   

Abstract

Modern hospital sterile service departments (SSDs) routinely inspect instruments visually to assess the degree of contamination present after the washer/disinfector cycle. We aimed to test the effectiveness and reliability of this method. Surgical instrument sets were obtained from nine anonymous National Health Service (NHS) Primary Care Trust SSDs to investigate the efficacy of 'in-place' cleaning procedures. The instruments were first inspected visually, followed by a novel technique called episcopic differential interference contrast microscopy. This was combined with a sensitive fluorescent reagent, SYPRO Ruby, to rapidly visualise and assess contamination levels on the medical devices. The application of a Contamination Index (0-4) for both proteinaceous and non-proteinaceous deposits on the surface allowed quantitative assessment. Close correlation was seen for simple instruments between visual assessment and microscopic analysis. For more complex instruments, however, there was a marked difference between the two assessment techniques and the microscopy procedure showed areas of proteinaceous and non-proteinaceous crystalline soiling that was difficult or impossible to see by eye. Visual assessment of cleaning is fraught with possible error. This survey shows how large amounts of contamination could pass undetected using such antiquated methods. The new methodology applied in the assessment of surface contamination is rapid and generally applicable and could be used more widely for routine monitoring of instrument cleanliness.

Mesh:

Substances:

Year:  2007        PMID: 17942186     DOI: 10.1016/j.jhin.2007.08.009

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  2 in total

1.  Efficacy of soaking in 70% isopropyl alcohol on aerobic bacterial decontamination of surgical instruments and gloves for serial mouse laparotomies.

Authors:  Jessica N Keen; MaryKay Austin; Li-Shan Huang; Susan Messing; Jeffrey D Wyatt
Journal:  J Am Assoc Lab Anim Sci       Date:  2010-11       Impact factor: 1.232

2.  Multicenter comparative study of three "non-destructive" methods of detecting the cleanliness of the da Vinci surgical robotic instrument.

Authors:  Aiqin Chen; Xiumei Zou; Yanhua Tan; Yanhan Chen; Xinru Ye; Shuqin Hao
Journal:  Gland Surg       Date:  2021-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.