A computer-controlled ultrasound pulser-receiver system for transskull fluid detection using a shear wave transmission technique.

Sai Chun Tang1, Gregory T Clement, Kullervo Hynynen.   

Abstract

The purpose of this study was to evaluate the performance of a computer-controlled ultrasound pulser-receiver system incorporating a shear mode technique for transskull fluid detection. The presence of fluid in the sinuses of an ex vivo human skull was examined using a pulse-echo method by transmitting an ultrasound beam through the maxilla bone toward the back wall on the other side of the sinus cavity. The pulser was programmed to generate bipolar pulse trains with 5 cycles at a frequency of 1 MHz, repetition frequency of about 20 Hz, and amplitude of 100 V to drive a 1-MHz piezoelectric transducer. Shear and longitudinal waves in the maxilla bone were produced by adjusting the bone surface incident angle to 45 degrees and 0 degrees, respectively. Computer tomography (CT) scans of the skull were performed to verify the ultrasound experiment. Using the shear mode technique, the echo waveform clearly distinguishes the presence of fluid, and the estimated distance of the ultrasound traveled in the sinus is consistent with the measurement from the CT images. Contrarily, using the longitudinal mode, no detectable back wall echo was observed under the same conditions. As a conclusion, this study demonstrated that the proposed pulser-receiver system with the shear mode technique is promising for transskull fluid detecting, such as mucus in a sinus.

Entities:  

Mesh:

Year:  2007        PMID: 17941383      PMCID: PMC2186207          DOI: 10.1109/tuffc.2007.461

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  18 in total

1.  Antimicrobial treatment guidelines for acute bacterial rhinosinusitis. Executive summary. Sinus and Allergy Health Partnership.

Authors: 
Journal:  Otolaryngol Head Neck Surg       Date:  2000-07       Impact factor: 3.497

Review 2.  Endodontic implications of the maxillary sinus: a review.

Authors:  C H J Hauman; N P Chandler; D C Tong
Journal:  Int Endod J       Date:  2002-02       Impact factor: 5.264

3.  Comparison of ultrasound, radiography, and clinical examination in the diagnosis of acute maxillary sinusitis: a systematic review.

Authors:  H Varonen; M Mäkelä; S Savolainen; E Läärä; J Hilden
Journal:  J Clin Epidemiol       Date:  2000-09       Impact factor: 6.437

4.  Enhanced ultrasound transmission through the human skull using shear mode conversion.

Authors:  G T Clement; P J White; K Hynynen
Journal:  J Acoust Soc Am       Date:  2004-03       Impact factor: 1.840

Review 5.  Accurate diagnosis and appropriate treatment of acute bacterial rhinosinusitis: minimizing bacterial resistance.

Authors:  Javier Garau; Ron Dagan
Journal:  Clin Ther       Date:  2003-07       Impact factor: 3.393

6.  Principles of appropriate antibiotic use for treatment of acute respiratory tract infections in adults: background, specific aims, and methods.

Authors:  R Gonzales; J G Bartlett; R E Besser; R J Cooper; J M Hickner; J R Hoffman; M A Sande
Journal:  Ann Intern Med       Date:  2001-03-20       Impact factor: 25.391

7.  Validity of ultrasonography in diagnosis of acute maxillary sinusitis.

Authors:  T Puhakka; T Heikkinen; M J Mäkelä; A Alanen; T Kallio; L Korsoff; J Suonpää; O Ruuskanen
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2000-12

Review 8.  Antimicrobial treatment guidelines for acute bacterial rhinosinusitis.

Authors:  Jack B Anon; Michael R Jacobs; Michael D Poole; Paul G Ambrose; Mark S Benninger; James A Hadley; William A Craig
Journal:  Otolaryngol Head Neck Surg       Date:  2004-01       Impact factor: 3.497

Review 9.  The clinical diagnosis of acute purulent sinusitis in general practice--a review.

Authors:  Morten Lindbaek; Per Hjortdahl
Journal:  Br J Gen Pract       Date:  2002-06       Impact factor: 5.386

Review 10.  Rhinosinusitis.

Authors:  Welby Winstead
Journal:  Prim Care       Date:  2003-03       Impact factor: 2.907

View more
  3 in total

1.  A PVDF receiver for ultrasound monitoring of transcranial focused ultrasound therapy.

Authors:  Meaghan A O'Reilly; Kullervo Hynynen
Journal:  IEEE Trans Biomed Eng       Date:  2010-05-27       Impact factor: 4.538

2.  An intraoperative brain shift monitor using shear mode transcranial ultrasound: preliminary results.

Authors:  P Jason White; Stephen Whalen; Sai Chun Tang; Greg T Clement; Ferenc Jolesz; Alexandra J Golby
Journal:  J Ultrasound Med       Date:  2009-02       Impact factor: 2.153

Review 3.  Driving circuitry for focused ultrasound noninvasive surgery and drug delivery applications.

Authors:  Munir M El-Desouki; Kullervo Hynynen
Journal:  Sensors (Basel)       Date:  2011-01-07       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.