Dosimetric characteristics of electron beams produced by a mobile accelerator for IORT.

M Pimpinella1, D Mihailescu, A S Guerra, R F Laitano.   

Abstract

Energy and angular distributions of electron beams with different energies were simulated by Monte Carlo calculations. These beams were generated by the NOVAC7 system (Hitesys, Italy), a mobile electron accelerator specifically dedicated to intra-operative radiation therapy (IORT). The electron beam simulations were verified by comparing the measured dose distributions with the corresponding calculated distributions. As expected, a considerable difference was observed in the energy and angular distributions between the IORT beams studied in the present work and the electron beams produced by conventional accelerators for non-IORT applications. It was also found that significant differences exist between the IORT beams used in this work and other IORT beams with different collimation systems. For example, the contribution from the scattered electrons to the total dose was found to be up to 15% higher in the NOVAC7 beams. The water-to-air stopping power ratios of the IORT beams used in this work were calculated on the basis of the beam energy distributions obtained by the Monte Carlo simulations. These calculated stopping power ratios, s(w,air), were compared with the corresponding s(w,air) values recommended by the TRS-381 and TRS-398 IAEA dosimetry protocols in order to estimate the deviations between a dosimetry based on generic parameters and a dosimetry based on parameters specifically obtained for the actual IORT beams. The deviations in the s(w,air) values were found to be as large as up to about 1%. Therefore, we recommend that a preliminary analysis should always be made when dealing with IORT beams in order to assess to what extent the possible differences in the s(w,air) values have to be accounted for or may be neglected on the basis of the specific accuracy needed in clinical dosimetry.

Entities:  

Mesh:

Year:  2007        PMID: 17921580     DOI: 10.1088/0031-9155/52/20/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Monte Carlo Calculation of the Energy Spectrum of a 6 MeV Electron Beam using PENetration and Energy Loss of Positrons and Electrons Code.

Authors:  Danny Giancarlo Apaza Veliz; Jorge Homero Wilches Visbal; Felipe Chen Abrego; José Luis Vega Ramírez
Journal:  J Med Phys       Date:  2020-07-20

Review 2.  Present state and issues in IORT Physics.

Authors:  Frank W Hensley
Journal:  Radiat Oncol       Date:  2017-01-27       Impact factor: 3.481

3.  Monte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator.

Authors:  M Robatjazi; K Tanha; S R Mahdavi; H R Baghani; H R Mirzaei; M Mousavi; N Nafissi; E Akbari
Journal:  J Biomed Phys Eng       Date:  2018-03-01

4.  Comparing the dosimetric characteristics of the electron beam from dedicated intraoperative and conventional radiotherapy accelerators.

Authors:  Hamid Reza Baghani; Seyed Mahmoud Reza Aghamiri; Seyed Rabi Mahdavi; Mohammad Esmail Akbari; Hamid Reza Mirzaei
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

5.  Dosimetric characteristics of electron beams produced by two mobile accelerators, Novac7 and Liac, for intraoperative radiation therapy through Monte Carlo simulation.

Authors:  Sergio Righi; Evis Karaj; Giuseppe Felici; Fabio Di Martino
Journal:  J Appl Clin Med Phys       Date:  2013-01-07       Impact factor: 2.102

6.  Dose distribution changes with shielding disc misalignments and wrong orientations in breast IOERT: a Monte Carlo - GEANT4 and experimental study.

Authors:  Giorgio Russo; Carlo Casarino; Gaetano Arnetta; Giuliana Candiano; Alessandro Stefano; Filippo Alongi; Giovanni Borasi; Cristina Messa; Maria C Gilardi
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.