Electrocardiographic optimization of interventricular delay in cardiac resynchronization therapy: a simple method to optimize the device.

Bàrbara Vidal1, David Tamborero, Lluis Mont, Marta Sitges, Victoria Delgado, Antonio Berruezo, Ernesto Díaz-Infante, Jose M Tolosana, Carles Paré, Josep Brugada.   

Abstract

INTRODUCTION: Echocardiography is widely used to optimize CRT programming, but it is time-consuming. This study aimed to correlate the optimal interventricular pacing (V-V) interval obtained by echo with the optimal V-V interval obtained by a simpler method based on the surface ECG. METHODS AND
RESULTS: Three V-V intervals were tested: LV preactivation at -30 ms, simultaneous biventricular pacing (0 ms), and RV preactivation at +30 ms. The one that achieved the best LV synchrony was chosen as the optimal V-V. This result was then compared with two different ECG measurements. The first ECG method considered the best V-V to be that which achieved the narrowest QRS. The second V-V method consisted in measuring the interval from the pacing spike to the beginning of the fast deflexion of the QRS complex in leads V1, V2, first pacing from the LV (T1), and after from the RV (T2). The T2-T1 interval was considered as a surrogate measurement of interventricular delay and defined as the best V-V. A cohort of 31 consecutive patients treated with CRT was studied. Optimal V-V interval obtained by echo was -30 ms in 25 patients (80%), +30 ms in three patients (10%), and 0 ms in the remaining three patients (10%). Echo results had 32% coincidence with the first ECG method (r = 0.2, P = NS) and 83% coincidence with the second ECG method (r = 0.81 P< 0.001).
CONCLUSIONS: The time difference in the fast ventricular depolarization observed between RV and LV stimulation in the surface ECG shows a good correlation with the V-V optimization chosen according to echo.

Entities:  

Mesh:

Year:  2007        PMID: 17916140     DOI: 10.1111/j.1540-8167.2007.00983.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  10 in total

1.  Measurement precision in the optimization of cardiac resynchronization therapy.

Authors:  Robert G Turcott; Ronald M Witteles; Paul J Wang; Randall H Vagelos; Michael B Fowler; Euan A Ashley
Journal:  Circ Heart Fail       Date:  2010-02-22       Impact factor: 8.790

2.  [Optimized programming during cardiac resynchronization therapy].

Authors:  T Deneke; B Lemke; M Horlitz; A Drüke; A Mügge; P H Grewe; T Lawo
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2008-03

3.  Optimization of cardiac resynchronization therapy after implantation.

Authors:  Ayesha Hasan; William T Abraham
Journal:  Curr Treat Options Cardiovasc Med       Date:  2008-08

4.  [Optimized CRT programming: relevance and practical application].

Authors:  F Bode; F Schütte
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-09

5.  Echocardiography versus intracardiac electrocardiography-based optimization for cardiac resynchronization therapy : a comparative clinical long-term trial.

Authors:  C J Jensen; A Liadski; M Bell; C K Naber; O Bruder; G V Sabin; B Küpper; H Wieneke
Journal:  Herz       Date:  2011-10       Impact factor: 1.443

6.  Lack of clinical predictors of optimal V-V delay in patients with cardiac resynchronization devices.

Authors:  Avi Fischer; Riple Hansalia; Samantha Buckley; Robin Goldberg; Martin Goldman; Paul Muntner; Davendra Mehta; W Lane Duvall
Journal:  J Interv Card Electrophysiol       Date:  2009-01-16       Impact factor: 1.900

Review 7.  An Overview of Current Cardiac Resynchronization Therapy.

Authors:  Chien-Ming Cheng; Jin-Long Huang; Tsu-Juey Wu; Yu-Cheng Hsieh; Kuo-Yang Wang; Shih-Ann Chen
Journal:  Acta Cardiol Sin       Date:  2013-11       Impact factor: 2.672

8.  Electrocardiographic markers of cardiac resynchronization therapy response: delayed time to intrinsicoid deflection onset in lateral leads.

Authors:  Rubén Ka Tapia-Orihuela; S Michael Gharacholou; Samuel J Asirvatham; Freddy Del-Carpio Munoz
Journal:  J Geriatr Cardiol       Date:  2022-01-28       Impact factor: 3.327

9.  A study of mechanical optimization strategy for cardiac resynchronization therapy based on an electromechanical model.

Authors:  Jianhong Dou; Ling Xia; Dongdong Deng; Yunliang Zang; Guofa Shou; Cesar Bustos; Weifeng Tu; Feng Liu; Stuart Crozier
Journal:  Comput Math Methods Med       Date:  2012-10-16       Impact factor: 2.238

10.  Electrocardiographic patterns in biventricular pacing delivered by second-generation cardiac resynchronization devices.

Authors:  Amirfarjam Fazelifar; Fatemeh Jorfi; Majid Haghjoo
Journal:  Indian Pacing Electrophysiol J       Date:  2017-11-04
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.