Reliability of brachial artery flow-mediated dilatation measurement using ultrasound.

Cláudia de Mello Meirelles1, Sandra Pereira Leite, Carlos Antonio Barbosa Montenegro, Paulo Sergio Chagas Gomes.   

Abstract

OBJECTIVE: To determine the reliability of the baseline (BBAB) and post occlusion (POBAD) brachial artery diameters, brachial artery flow-mediated dilatations (BAFMD) measurements, and to quantify the typical error of the measurements (TEM).
METHODS: Internal consistency (2 measurements on the same day) was determined in 10 volunteers, whereas stability (2 measurements on separate days) was determined in 13 volunteers. All the volunteers looked healthy and were nonsmokers. The brachial artery images were obtained using a two dimensional Doppler ultrasound instrument: a 14 MHz transducer was used. The distances between the intima-lumen interfaces were measured before and after the blood flow had been stopped by means of a cuff on the arm for five minutes. BAFMD was considered as the percentile increase of POBAD in relation to BBAD.
RESULTS: ANOVA did not identify any significant differences between the measurements taken on the same and separate days. For BAFMD, the intraclass correlation coefficients between the measurements taken on the same and separate days were: R = 0.7001 and R = 0.8420, respectively (p < 0.05). The variation coefficients were 5.8% and 12.4% and the relative SMEs were 13.8% and 14.9%, respectively, for the measurements taken on the same and separate days. Analysis of the Bland-Altman graphs indicated that the variables did not present heteroscedastic errors.
CONCLUSION: The BBAD, POBAD and BAFMD measurements using the manual ultrasound technique were highly reliable for both the same and separate day measurements and therefore can be used to diagnose and monitor endothelial function.

Entities:  

Mesh:

Year:  2007        PMID: 17906817     DOI: 10.1590/s0066-782x2007001500006

Source DB:  PubMed          Journal:  Arq Bras Cardiol        ISSN: 0066-782X            Impact factor:   2.000


  6 in total

Review 1.  Anatomical and Functional Estimations of Brachial Artery Diameter and Elasticity Using Oscillometric Measurements with a Quantitative Approach.

Authors:  Keiichiro Yoshinaga; Satoshi Fujii; Yuuki Tomiyama; Keisuke Takeuchi; Nagara Tamaki
Journal:  Pulse (Basel)       Date:  2016-03-05

2.  The effects of urate lowering therapy on inflammation, endothelial function, and blood pressure (SURPHER) study design and rationale.

Authors:  Michael B Saddekni; Kenneth G Saag; Tanja Dudenbostel; Suzanne Oparil; David A Calhoun; Sebastian E Sattui; Daniel I Feig; Paul Muntner; David T Redden; Phillip J Foster; Elizabeth J Rahn; Stephanie R Biggers; Peng Li; Angelo L Gaffo
Journal:  Contemp Clin Trials       Date:  2016-08-30       Impact factor: 2.226

3.  Effects of thyroxine replacement on endothelial function and carotid artery intima-media thickness in female patients with mild subclinical hypothyroidism.

Authors:  Monica Dias Cabral; Patricia Teixeira; Debora Soares; Sandra Leite; Elizabeth Salles; Mario Waisman
Journal:  Clinics (Sao Paulo)       Date:  2011       Impact factor: 2.365

4.  Accurate quantitative measurements of brachial artery cross-sectional vascular area and vascular volume elastic modulus using automated oscillometric measurements: comparison with brachial artery ultrasound.

Authors:  Yuuki Tomiyama; Keiichiro Yoshinaga; Satoshi Fujii; Noriki Ochi; Mamiko Inoue; Mutumi Nishida; Kumi Aziki; Tatsunori Horie; Chietsugu Katoh; Nagara Tamaki
Journal:  Hypertens Res       Date:  2015-02-19       Impact factor: 3.872

5.  Endothelial Effect of Statin Therapy at a High Dose Versus Low Dose Associated with Ezetimibe.

Authors:  Maristela Magnavita Oliveira Garcia; Carolina Garcez Varela; Patricia Fontes Silva; Paulo Roberto Passos Lima; Paulo Meira Góes; Marilia Galeffi Rodrigues; Maria de Lourdes Lima Souza E Silva; Ana Marice Teixeira Ladeia; Armênio Costa Guimarães; Luis Claudio Lemos Correia
Journal:  Arq Bras Cardiol       Date:  2016-04       Impact factor: 2.000

6.  Endothelial dysfunction assessment by flow-mediated dilation in a high-altitude population.

Authors:  Walter S Calderón-Gerstein; Antonio López-Peña; Raúl Macha-Ramírez; Astrid Bruno-Huamán; Roxana Espejo-Ramos; Stephany Vílchez-Bravo; María Ramírez-Breña; Milagros Damián-Mucha; Adriana Matos-Mucha
Journal:  Vasc Health Risk Manag       Date:  2017-11-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.