Rectangular cymbal arrays for improved ultrasonic transdermal insulin delivery.

Joseph Luis1, Eun Joo Park, Richard J Meyer, Nadine Barrie Smith.   

Abstract

Circular cymbal ultrasound arrays have been shown to be effective in delivering therapeutic levels of insulin in rats, rabbits, and pigs. To improve delivery efficiency, a rectangular cymbal design was desired in order to achieve a broader spatial intensity field without increasing the size of the device or the spatial-peak temporal-peak intensity (I(SPTP)). With a similar intensity (50 mWcm(2)), the goal was to determine if the 3x1 rectangular cymbal array could perform significantly better than the 3x3 circular array for glucose reduction in hyperglycemic rabbits. Rabbit experiments were performed using three groups: nonsonicated control (n=3), ultrasound exposure using a circular cymbal array (n=3), and ultrasound exposure using a rectangular cymbal array (n=3). Rabbits were anesthetized and a water tight reservoir that held the insulin was fastened on the rabbit's thigh. At the beginning of the experiment and every 15 min for 90 min, the blood glucose level was determined. For comparison between individual rabbits, the absolute level is normalized by subtracting out the baseline in order to arrive at the change in glucose level. For the control group, the normalized glucose level increased (more hyperglycemic) to +80.0+/-28.8 mgdl (mean+/-SEM). Using the circular array, the glucose level decreased to -146.7+/-17.8 mgdl at 90 min. However, using the rectangular cymbal array, the glucose decreased faster and to a level of -200.8+/-5.9 mgdl after 90 min. These results indicated the feasibility of the rectangular cymbal array as an improved device for drug delivery.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17902839     DOI: 10.1121/1.2769980

Source DB:  PubMed          Journal:  J Acoust Soc Am        ISSN: 0001-4966            Impact factor:   1.840


  7 in total

Review 1.  Transdermal delivery of proteins.

Authors:  Haripriya Kalluri; Ajay K Banga
Journal:  AAPS PharmSciTech       Date:  2011-03-03       Impact factor: 3.246

Review 2.  Ultrasound-mediated transdermal drug delivery: mechanisms, scope, and emerging trends.

Authors:  Baris E Polat; Douglas Hart; Robert Langer; Daniel Blankschtein
Journal:  J Control Release       Date:  2011-01-14       Impact factor: 9.776

Review 3.  Low-frequency sonophoresis: application to the transdermal delivery of macromolecules and hydrophilic drugs.

Authors:  Baris E Polat; Daniel Blankschtein; Robert Langer
Journal:  Expert Opin Drug Deliv       Date:  2010-12       Impact factor: 6.648

Review 4.  Advances in transdermal insulin delivery.

Authors:  Yuqi Zhang; Jicheng Yu; Anna R Kahkoska; Jinqiang Wang; John B Buse; Zhen Gu
Journal:  Adv Drug Deliv Rev       Date:  2018-12-08       Impact factor: 15.470

5.  Dose comparison of ultrasonic transdermal insulin delivery to subcutaneous insulin injection.

Authors:  Eun-Joo Park; Jeff Dodds; Nadine Barrie Smith
Journal:  Int J Nanomedicine       Date:  2008

6.  Enhanced Skin Permeation of Anti-wrinkle Peptides via Molecular Modification.

Authors:  Seng Han Lim; Yuanyuan Sun; Thulasi Thiruvallur Madanagopal; Vinicius Rosa; Lifeng Kang
Journal:  Sci Rep       Date:  2018-01-25       Impact factor: 4.379

Review 7.  Novel Antiretroviral Therapeutic Strategies for HIV.

Authors:  Rita F Cunha; Sandra Simões; Manuela Carvalheiro; José M Azevedo Pereira; Quirina Costa; Andreia Ascenso
Journal:  Molecules       Date:  2021-08-31       Impact factor: 4.411

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.