Literature DB >> 17898024

Exit of catheter lock solutions from double lumen acute haemodialysis catheters--an in vitro study.

Murat Sungur1, Emel Eryuksel, Sinan Yavas, Azra Bihorac, A Joseph Layon, Lawrence Caruso.   

Abstract

BACKGROUND: Double lumen dialysis catheters are routinely heparin or citrate 'locked' to maintain patency. Heparin lock-related bleeding episodes and antibiotic lock-related toxicity have been reported. The aim of this study is to quantify the amount of leak during 'lock' procedures and to compare leakage for different double lumen dialysis catheters.
METHODS: In an experimental, in vitro study at a University research laboratory, five different double lumen dialysis catheters were tested using three different lock volumes.
RESULTS: Using the catheter flush volume, leak ratios for Flexxicon II 15 cm and 20 cm catheters were greater than that seen in the Arrow 16 cm catheter (P < 0.05). Using 20% less than the catheter flush volume, the Flexxicon II 20 cm catheter had greater leak than the Duo-flow 15 and 20 cm catheters and Arrow 16 cm catheter (P < 0.05). The Flexxicon II 15 cm catheter had greater leak than the Duo-flow 15 cm and Duo-flow 20 cm catheters with 20% less locking volume (P < 0.05). Using 20% greater than the catheter flush volume, the Duo-flow 20 cm catheter had significantly less leak ratio than the Flexxicon II 20 cm catheter (P < 0.05). There were no other significant differences in leak ratios between the catheters.
CONCLUSION: All double lumen dialysis catheters we tested have a substantial amount of leak even when the catheter 'lock' volumes were used, and leak ratio increases significantly with 20% overfill. There is a leak even when using 20% less 'lock' volume. The amount of leak can be clinically important and may explain the reports of bleeding episodes after heparin lock and antibiotic toxicity after antibiotic and anticoagulant combination lock. Some devices have lower leak ratios than others, likely related to catheter design.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17898024     DOI: 10.1093/ndt/gfm452

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  9 in total

1.  Quasi-experimental study of sodium citrate locks and the risk of acute hemodialysis catheter infection among critically ill patients.

Authors:  Jean-Jacques Parienti; Stéphanie Deryckère; Bruno Mégarbane; Xavier Valette; Amélie Seguin; Bertrand Sauneuf; Jean-Paul Mira; Bertrand Souweine; Vincent Cattoir; Cédric Daubin; Damien du Cheyron
Journal:  Antimicrob Agents Chemother       Date:  2014-06-30       Impact factor: 5.191

Review 2.  Current state of the art for renal replacement therapy in critically ill patients with acute kidney injury.

Authors:  Sean M Bagshaw; Michael Darmon; Marlies Ostermann; Fredric O Finkelstein; Ron Wald; Ashita J Tolwani; Stuart L Goldstein; David J Gattas; Shigehiko Uchino; Eric A Hoste; Stephane Gaudry
Journal:  Intensive Care Med       Date:  2017-03-13       Impact factor: 17.440

3.  Does the heparin lock concentration affect hemodialysis catheter patency?

Authors:  D Maya Ivan; Tamorie Smith; Michael Allon
Journal:  Clin J Am Soc Nephrol       Date:  2010-05-24       Impact factor: 8.237

4.  Comparison of complications and dialysis adequacy between temporary and permanent tunnelled catheter for haemodialysis.

Authors:  Nihad Kukavica; Halima Resic; Vahidin Sahovic
Journal:  Bosn J Basic Med Sci       Date:  2009-11       Impact factor: 3.363

5.  Prevention of catheter-related bacteremia in children on hemodialysis: time for action.

Authors:  Constantinos J Stefanidis
Journal:  Pediatr Nephrol       Date:  2009-07-23       Impact factor: 3.714

6.  Convective Leakage Makes Heparin Locking of Central Venous Catheters Ineffective Within Seconds: Experimental Measurements in a Model Superior Vena Cava.

Authors:  Michael C Barbour; Patrick M McGah; Chin H Ng; Alicia M Clark; Kenneth W Gow; Alberto Aliseda
Journal:  ASAIO J       Date:  2015 Nov-Dec       Impact factor: 2.872

7.  Minimizing hemodialysis catheter dysfunction: an ounce of prevention.

Authors:  Timmy Lee; Charmaine Lok; Miguel Vazquez; Louise Moist; Ivan Maya; Michele Mokrzycki
Journal:  Int J Nephrol       Date:  2012-02-19

Review 8.  Reevaluation of lock solutions for Central venous catheters in hemodialysis: a narrative review.

Authors:  Yiqin Wang; Xuefeng Sun
Journal:  Ren Fail       Date:  2022-12       Impact factor: 3.222

9.  Use of tissue plasminogen activator in catheters used for extracorporeal renal replacement therapy.

Authors:  C Langston; A Eatroff; K Poeppel
Journal:  J Vet Intern Med       Date:  2014-01-17       Impact factor: 3.333

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.