Pacemaker lead prolapse through the pulmonary valve in children.

Charles I Berul1, Juan Villafane, Dianne L Atkins, Frank Cecchin, Joel A Kirsh, James A Johns, Ronald J Kanter, Peter P Karpawich.   

Abstract

BACKGROUND: Transvenous pacemaker leads in children are often placed with redundant lead length to allow for anticipated patient growth. This excess lead may rarely prolapse into the pulmonary artery and potentially interfere with valve function. We sought to determine the response to lead repositioning on pulmonary valve insufficiency.
METHODS: Retrospective reviews of demographics, lead type, implant duration, and radiography and echocardiography.
RESULTS: A total of 11 pediatric patients were identified with lead prolapse through the pulmonary valve, of which nine patients underwent procedures to retract and reposition the lead (age at implant 9 +/- 4 years, age at revision 13 +/- 4 years). The implant duration prior to revision was 4 +/- 3 years. Two leads required radiofrequency extraction sheaths for removal, two pulled back using a snare, while five leads were simply retracted and repositioned. Tricuspid regurgitation was none/trivial (three), mild (four), or moderate (two) and only two improved with repositioning or replacement. Pulmonary regurgitation preoperatively was mild (three), mild-moderate (two), or moderate (four) compared with trivial (three), mild (four), and moderate (two) after revision. Patients with longer-term implanted leads had less improvement in pulmonary insufficiency. Two patients had mild pulmonary stenosis from lead-related obstruction.
CONCLUSIONS: Prolapse of transvenous pacing leads into the pulmonary artery can occur when excess slack is left for growth. Leads can often be repositioned, but may require extraction and replacement, particularly if chronically implanted and adherent to valve apparatus. Lead revision does not always resolve pulmonary insufficiency, potentially leaving permanent valve damage.

Entities:  

Mesh:

Year:  2007        PMID: 17897120     DOI: 10.1111/j.1540-8159.2007.00839.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Severe tricuspid valve stenosis secondary to pacemaker leads presenting as ascites and liver dysfunction: a complex problem requiring a multidisciplinary therapeutic approach.

Authors:  Anita Krishnan; Achintya Moulick; Pranava Sinha; Karen Kuehl; Joshua Kanter; Michael Slack; Jonathan Kaltman; Marco Mercader; Jeffrey P Moak
Journal:  J Interv Card Electrophysiol       Date:  2008-10-11       Impact factor: 1.900

2.  2021 PACES expert consensus statement on the indications and management of cardiovascular implantable electronic devices in pediatric patients.

Authors:  Maully J Shah; Michael J Silka; Jennifer N Avari Silva; Seshadri Balaji; Cheyenne M Beach; Monica N Benjamin; Charles I Berul; Bryan Cannon; Frank Cecchin; Mitchell I Cohen; Aarti S Dalal; Brynn E Dechert; Anne Foster; Roman Gebauer; M Cecilia Gonzalez Corcia; Prince J Kannankeril; Peter P Karpawich; Jeffery J Kim; Mani Ram Krishna; Peter Kubuš; Martin J LaPage; Douglas Y Mah; Lindsey Malloy-Walton; Aya Miyazaki; Kara S Motonaga; Mary C Niu; Melissa Olen; Thomas Paul; Eric Rosenthal; Elizabeth V Saarel; Massimo Stefano Silvetti; Elizabeth A Stephenson; Reina B Tan; John Triedman; Nicholas H Von Bergen; Philip L Wackel
Journal:  Indian Pacing Electrophysiol J       Date:  2021-07-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.