| Literature DB >> 17896022 |
Wee-Liat Ong1, Kum-Cheong Tang, Ajay Agarwal, Ranganathan Nagarajan, Lian-Wee Luo, Levent Yobas.
Abstract
High-throughput screening of drug candidates for channelopathies can greatly benefit from an automated patch-clamping assay. Automation of the patch clamping through microfluidics ideally requires on-chip integration of glass capillaries with substantially round cross section. Such round capillaries, if they can only be integrated to connect isolated reservoirs on a substrate surface, will lead to a "lateral" configuration which is simple yet powerful for the patch clamping. We demonstrate here "lateral" patch clamping through microfluidic integration of substantially round glass capillaries in a novel process. The process adopts two well-known phenomena from microelectronics: keyhole-void formation and thermal-reflow of phosphosilicate glass in silicon trenches. The process relies on the same physical principle as the preparation of conventional micropipette electrodes by heat-pulling and fire-polishing glass tubes. The optimized process forms capillaries with a diameter approximately 1.5 microm and variation <10%. Functionality of the integrated glass capillaries for the patch-clamp recording has been verified by statistical test results from a sample of one hundred capillaries on mammalian cells (RBL-1) in suspension: 61% formed gigaseals (>1 GOmega) and of those approximately 48% (29% of all) achieved whole-cell recordings. Pharmacological blockade of ion channel activity and longevity of a whole-cell mode on these capillaries have also been presented.Entities:
Mesh:
Year: 2007 PMID: 17896022 DOI: 10.1039/b707439e
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799