Comparison between pulse oximetry and transthoracic impedance alarm traces during home monitoring.

N Nassi1, R Piumelli, E Lombardi, L Landini, G Donzelli, M de Martino.   

Abstract

OBJECTIVE: To compare transthoracic impedance (TTI/ECG) and pulse oximetry alarm traces detected during home monitoring in infants at risk of apnoea, bradycardia and hypoxaemia. STUDY
DESIGN: A retrospective evaluation of the monitor downloads of 67 infants who had undergone either TTI/ECG or pulse oximetry home monitoring using a device which can detect both parameters.
METHODS: The patients were categorised as: apparent life-threatening events (n = 39), preterm infants (n = 21) and miscellaneous (n = 7). TTI/ECG and pulse oximetry alarm traces were scored as either true or false alarms. Classification criteria were based on visual analysis of the impedance and plethysmographic waveforms captured by the memory monitor every time alarm thresholds were violated.
RESULTS: 5242 alarms occurred over 3452 days of monitoring: 4562 (87%) were false and 680 (13%) true. The mean duration of monitoring was 51 days (range 5-220 days). There were 2982 TTI/ECG false alarms (65% of the total) and 1580 pulse oximetry false alarms (35%) (p = 0.0042). Of the 680 true alarms, 507 (74%) were desaturations not attributable to central apnoea and 173 (26%) were true TTI/ECG alarms (p = 0.0013).
CONCLUSIONS: Comparison of pulse oximetry and TTI/ECG alarm traces shows that true events were mostly attributable to desaturations, while false alarms were mainly provoked by TTI/ECG. The total number of false alarms is lower than reported in other studies using TTI/ECG only, thus indicating that monitoring using both pulse oximetry and TTI/ECG is suitable for home use. When the combination of both techniques is not feasible or not required, we recommend the use of motion resistant pulse oximetry alone.

Entities:  

Mesh:

Year:  2007        PMID: 17893118     DOI: 10.1136/adc.2007.118513

Source DB:  PubMed          Journal:  Arch Dis Child        ISSN: 0003-9888            Impact factor:   3.791


  5 in total

1.  Long-term tolerability of capnography and respiratory inductance plethysmography for respiratory monitoring in pediatric patients treated with patient-controlled analgesia.

Authors:  Karen M Miller; Andrew Y Kim; Myron Yaster; Sapna R Kudchadkar; Elizabeth White; James Fackler; Constance L Monitto
Journal:  Paediatr Anaesth       Date:  2015-06-03       Impact factor: 2.556

2.  Influences of environmental noise level and respiration rate on the accuracy of acoustic respiration rate monitoring.

Authors:  Shizuha Yabuki; Hiroaki Toyama; Yusuke Takei; Toshihiro Wagatsuma; Hiroshi Yabuki; Masanori Yamauchi
Journal:  J Clin Monit Comput       Date:  2017-02-07       Impact factor: 2.502

3.  Linshom respiratory monitoring device: a novel temperature-based respiratory monitor.

Authors:  Jerrold Lerman; Doron Feldman; Ronen Feldman; John Moser; Leeshi Feldman; Madhankumar Sathyamoorthy; Kenneth Deitch; Uri Feldman
Journal:  Can J Anaesth       Date:  2016-07-13       Impact factor: 6.713

Review 4.  Apparent Life-Threatening Events (ALTE): Italian guidelines.

Authors:  Raffaele Piumelli; Riccardo Davanzo; Niccolò Nassi; Silvia Salvatore; Cinzia Arzilli; Marta Peruzzi; Massimo Agosti; Antonella Palmieri; Maria Giovanna Paglietti; Luana Nosetti; Raffaele Pomo; Francesco De Luca; Alessandro Rimini; Salvatore De Masi; Simona Costabel; Valeria Cavarretta; Anna Cremante; Fabio Cardinale; Renato Cutrera
Journal:  Ital J Pediatr       Date:  2017-12-12       Impact factor: 2.638

5.  Electrical Bioimpedance Spectroscopy on Acute Unilateral Stroke Patients: Initial Observations regarding Differences between Sides.

Authors:  Fernando Seoane; Seyed Reza Atefi; Jens Tomner; Konstantinos Kostulas; Kaj Lindecrantz
Journal:  Biomed Res Int       Date:  2015-10-18       Impact factor: 3.411

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.