BACKGROUND: Accurate and reproducible measurement of left ventricular (LV) mass and function is a significant strength of Cardiovascular Magnetic Resonance (CMR). Reproducibility and accuracy of these measurements is usually reported between experienced operators. However, an increasing number of inexperienced operators are now training in CMR and are involved in post-processing analysis. The aim of the study was to assess the interobserver variability of the manual planimetry of LV contours amongst two experienced and six inexperienced operators before and after a two months training period.
METHODS: Ten healthy normal volunteers (5 men, mean age 34+/-14 years) comprised the study population. LV volumes, mass, and ejection fraction were manually evaluated using Argus software (Siemens Medical Solutions, Erlangen, Germany) for each subject, once by the two experienced and twice by the six inexperienced operators. The mean values of experienced operators were considered the reference values. The agreement between operators was evaluated by means of Bland-Altman analysis. Training involved standardized data acquisition, simulated off-line analysis and mentoring.
RESULTS: The trainee operators demonstrated improvement in the measurement of all the parameters compared to the experienced operators. The mean ejection fraction variability improved from 7.2% before training to 3.7% after training (p=0.03). The parameter in which the trainees showed the least improvement was LV mass (from 7.7% to 6.7% after training). The basal slice selection and contour definition were the main sources of errors.
CONCLUSIONS: An intensive two month training period significantly improved the accuracy of LV functional measurements. Adequate training of new CMR operators is of paramount importance in our aim to maintain the accuracy and high reproducibility of CMR in LV function analysis.

Entities:  

Mesh:

Year:  2007        PMID: 17891615     DOI: 10.1080/10976640701545073

Source DB:  PubMed          Journal:  J Cardiovasc Magn Reson        ISSN: 1097-6647            Impact factor:   5.364


  51 in total

Review 1.  Novel techniques for assessment of left ventricular systolic function.

Authors:  Sonal Chandra; Hicham Skali; Ron Blankstein
Journal:  Heart Fail Rev       Date:  2011-07       Impact factor: 4.214

Review 2.  Cardiovascular magnetic resonance in heart failure.

Authors:  Theodoros D Karamitsos; Stefan Neubauer
Journal:  Curr Cardiol Rep       Date:  2011-06       Impact factor: 2.931

Review 3.  Evaluation of left ventricular function using cardiac magnetic resonance imaging.

Authors:  Suchi Grover; Darryl P Leong; Joseph B Selvanayagam
Journal:  J Nucl Cardiol       Date:  2011-04       Impact factor: 5.952

4.  Diagnostic accuracy and variability of three semi-quantitative methods for assessing right ventricular systolic function from cardiac MRI in patients with acquired heart disease.

Authors:  Jérôme Caudron; Jeannette Fares; Pierre-Hugues Vivier; Valentin Lefebvre; Caroline Petitjean; Jean-Nicolas Dacher
Journal:  Eur Radiol       Date:  2011-05-26       Impact factor: 5.315

Review 5.  Cardiac magnetic resonance imaging in heart failure: where the alphabet begins!

Authors:  Ahmed Aljizeeri; Abdulbaset Sulaiman; Naji Alhulaimi; Ahmed Alsaileek; Mouaz H Al-Mallah
Journal:  Heart Fail Rev       Date:  2017-07       Impact factor: 4.214

6.  Native T1 mapping to detect extent of acute and chronic myocardial infarction: comparison with late gadolinium enhancement technique.

Authors:  Amardeep Ghosh Dastidar; Iwan Harries; Giulia Pontecorboli; Vito D Bruno; Estefania De Garate; Charlie Moret; Anna Baritussio; Thomas W Johnson; Elisa McAlindon; Chiara Bucciarelli-Ducci
Journal:  Int J Cardiovasc Imaging       Date:  2018-10-24       Impact factor: 2.357

7.  Feasibility and safety of high-dose adenosine perfusion cardiovascular magnetic resonance.

Authors:  Theodoros D Karamitsos; Ntobeko A B Ntusi; Jane M Francis; Cameron J Holloway; Saul G Myerson; Stefan Neubauer
Journal:  J Cardiovasc Magn Reson       Date:  2010-11-16       Impact factor: 5.364

8.  A dual propagation contours technique for semi-automated assessment of systolic and diastolic cardiac function by CMR.

Authors:  Wei Feng; Hosakote Nagaraj; Himanshu Gupta; Steven G Lloyd; Inmaculada Aban; Gilbert J Perry; David A Calhoun; Louis J Dell'Italia; Thomas S Denney
Journal:  J Cardiovasc Magn Reson       Date:  2009-08-13       Impact factor: 5.364

9.  Recommendations for cardiovascular magnetic resonance in adults with congenital heart disease from the respective working groups of the European Society of Cardiology.

Authors:  Philip J Kilner; Tal Geva; Harald Kaemmerer; Pedro T Trindade; Juerg Schwitter; Gary D Webb
Journal:  Eur Heart J       Date:  2010-01-11       Impact factor: 29.983

10.  Intra-observer and interobserver variability of biventricular function, volumes and mass in patients with congenital heart disease measured by CMR imaging.

Authors:  Saskia E Luijnenburg; Daniëlle Robbers-Visser; Adriaan Moelker; Hubert W Vliegen; Barbara J M Mulder; Willem A Helbing
Journal:  Int J Cardiovasc Imaging       Date:  2009-09-16       Impact factor: 2.357

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.