Calibration errors of Goldmann tonometers in a tertiary eye care centre.

J Y Chuo1, F S Mikelberg.   

Abstract

BACKGROUND: To evaluate the calibration errors of all Goldmann applanation tonometers used clinically in the Vancouver General Hospital/University of British Columbia (VGH/UBC) Eye Care Centre.
METHODS: All Goldmann tonometers used for patient care in the VGH/UBC Eye Care Centre were individually assessed for calibration errors. The tonometers were checked according to the Haag-Streit method using a standard calibration weight bar at 3 pressure levels (0, 20, 60 mm Hg). Calibration errors were classified into 3 categories: +/-0.5 mm Hg, +/-1.0 mm Hg, or +/-1.5 mm Hg or more. Ophthalmologists using tonometers with calibration errors > +/-1.0 mm Hg were notified, with the suggestion that the tonometers be returned to the manufacturer for recalibration.
RESULTS: A total of 25 tonometers were evaluated. At the 20 mm Hg level, 17 (68%) fell within the manufacturer's recommended calibration range of +/-0.5 mm Hg, and 2 (8%) had errors of +/-1.0 mm Hg, resulting in 19 tonometers (76%) with errors of +/-1.0 mm Hg. Six tonometers (24%) were identified as having calibration errors greater than +/-1.0 mm Hg.
INTERPRETATION: Calibration error of +/-1.0 mm Hg or less is clinically acceptable. Seventy-six percent of the Goldmann applanation tonometers in the Eye Care Centre were in the acceptable calibration range. As the number of acceptably calibrated tonometers was high, despite the infrequency of recent calibration, we feel that monthly evaluation of Goldmann tonometers is suitable.

Entities:  

Mesh:

Year:  2007        PMID: 17891198     DOI: 10.3129/i07-142

Source DB:  PubMed          Journal:  Can J Ophthalmol        ISSN: 0008-4182            Impact factor:   1.882


  3 in total

1.  Potential effects of systematic errors in intraocular pressure measurements on screening for ocular hypertension.

Authors:  M J Turner; S L Graham; A P Avolio; P Mitchell
Journal:  Eye (Lond)       Date:  2013-02-22       Impact factor: 3.775

2.  Prevalence of calibration errors in goldmann applanation tonometers.

Authors:  Heydar Amini; Sasan Moghimi; Mahboobe Golestani; Mahmood Jabbarvand; Reza Zarei; Mohammadreza Esmaili; Zohre Esfandiarpour; Reza Soltani-Moghadam
Journal:  J Ophthalmic Vis Res       Date:  2009-07

3.  Rectifying calibration error of Goldmann applanation tonometer is easy!

Authors:  Nikhil S Choudhari; Krishna P Moorthy; Vinod B Tungikar; Mohan Kumar; Ronnie George; Harsha L Rao; Sirisha Senthil; Lingam Vijaya; Chandra Sekhar Garudadri
Journal:  Indian J Ophthalmol       Date:  2014-11       Impact factor: 1.848

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.