Validation of the HEM-780REL with easy wrap cuff for self-measurement of blood pressure according to the European Society of Hypertension International Protocol.

Anthony J Viera1, Alan L Hinderliter.   

Abstract

OBJECTIVE: We tested the accuracy of the HEM-780REL automatic blood pressure monitor with Easy Wrap Cuff.
METHODS: We used the International Protocol for validation of blood pressure measuring devices developed by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. Thirty-three adults over the age of 30 years participated to have 11 total participants in each of the three required blood pressure ranges for systolic and diastolic blood pressure. Sequential blood pressure readings were taken independently by trained observers using a mercury standard with appropriate size cuff. A third observer performed measurements with the test device. Analyses were performed according to International Protocol specifications for the 99 pairs of measurements. The device was given a pass/fail recommendation based on bands of accuracy compared with the mercury standard (within 5, 10, or 15 mmHg), as well as number of readings per participant within 5 mmHg.
RESULTS: The mean blood pressure difference was 0.52+/-7.7 mmHg for systolic blood pressure and 0.39+/-4.7 mmHg for diastolic blood pressure. Twenty-four out of 33 participants had two out of three readings within 5 mmHg of the mercury standard for systolic blood pressure. Twenty-seven out of 33 participants had two out of three readings within 5 mmHg of the mercury standard for diastolic blood pressure. The device received a passing grade both for accuracy of individual measurements and for accuracy for individuals.
CONCLUSION: The HEM-780REL with Easy Wrap Cuff performs accurately according to the standards of the International Protocol.

Entities:  

Mesh:

Year:  2007        PMID: 17890973     DOI: 10.1097/MBP.0b013e3281404a7e

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  5 in total

1.  Do clinicians tell patients they have prehypertension?

Authors:  Anthony J Viera; Fatima Bangura; C Madeline Mitchell; Ana Cerna; Philip Sloane
Journal:  J Am Board Fam Med       Date:  2011 Jan-Feb       Impact factor: 2.657

2.  Polysomnography-derived sleep parameters as a determinant of nocturnal blood pressure profile in patients with obstructive sleep apnea.

Authors:  Mitsuo Kuwabara; Naoko Tomitani; Toshikazu Shiga; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-05-30       Impact factor: 3.738

3.  Novel Triggered Nocturnal Blood Pressure Monitoring for Sleep Apnea Syndrome: Distribution and Reproducibility of Hypoxia-Triggered Nocturnal Blood Pressure Measurements.

Authors:  Mitsuo Kuwabara; Haruna Hamasaki; Naoko Tomitani; Toshikazu Shiga; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2016-07-14       Impact factor: 3.738

4.  Validation of the Oregon Scientific BPU 330 for self-monitoring of blood pressure according to the International Protocol.

Authors:  Li Li; XinYu Zhang; ChunHong Yan; QingXiang Liang
Journal:  Vasc Health Risk Manag       Date:  2008

5.  Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European Society of Hypertension International Protocol.

Authors:  Jirar Topouchian; Davide Agnoletti; Jacques Blacher; Ahmed Youssef; Mirna N Chahine; Isabel Ibanez; Nathalie Assemani; Roland Asmar
Journal:  Vasc Health Risk Manag       Date:  2014-01-16
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.