Reactive rise in blood pressure upon cuff inflation: cuff inflation at the arm causes a greater rise in pressure than at the wrist in hypertensive patients.

Alexia Charmoy1, Grégoire Würzner, Christiane Ruffieux, Christopher Hasler, François Cachat, Bernard Waeber, Michel Burnier.   

Abstract

OBJECTIVE: Cuff inflation at the arm is known to cause an instantaneous rise in blood pressure, which might be due to the discomfort of the procedure and might interfere with the precision of the blood pressure measurement. In this study, we compared the reactive rise in blood pressure induced by cuff inflation when the cuff was placed at the upper arm level and at the wrist. PARTICIPANTS AND METHODS: The reactive rise in systolic and diastolic blood pressure to cuff inflation was measured in 34 normotensive participants and 34 hypertensive patients. Each participant was equipped with two cuffs, one around the right upper arm (OMRON HEM-CR19, 22-32 cm) and one around the right wrist (OMRON HEM-CS 19, 17-22 cm; Omron Health Care Europe BV, Hoofddorp, The Netherlands). The cuffs were inflated in a double random order (maximal cuff pressure and position of the cuff) with two maximal cuff pressures: 180 and 240 mmHg. The cuffs were linked to an oscillometric device (OMRON HEM 907; Omron Health Care). Simultaneously, blood pressure was measured continuously at the middle finger of the left hand using photoplethysmography. Three measurements were made at each level of blood pressure at the arm and at the wrist, and the sequence of measurements was randomized.
RESULTS: In normotensive participants, no significant difference was observed in the reactive rise in blood pressure when the cuff was inflated either at the arm or at the wrist irrespective of the level of cuff inflation. Inflating a cuff at the arm, however, induced a significantly greater rise in blood pressure than inflating it at the wrist in hypertensive participants for both systolic and diastolic pressures (P<0.01), and at both levels of cuff inflation. The blood pressure response to cuff inflation was independent of baseline blood pressure.
CONCLUSIONS: The results show that in hypertensive patients, cuff inflation at the wrist produces a smaller reactive rise in blood pressure. The difference between the arm and the wrist is independent of the patient's level of blood pressure.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17890965     DOI: 10.1097/MBP.0b013e3282c9ac9a

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  10 in total

1.  Effect of vitamin D supplementation on blood pressure in blacks.

Authors:  John P Forman; Jamil B Scott; Kimmie Ng; Bettina F Drake; Elizabeth Gonzalez Suarez; Douglas L Hayden; Gary G Bennett; Paulette D Chandler; Bruce W Hollis; Karen M Emmons; Edward L Giovannucci; Charles S Fuchs; Andrew T Chan
Journal:  Hypertension       Date:  2013-04       Impact factor: 10.190

2.  Prevalence of pseudoresistant hypertension due to inaccurate blood pressure measurement.

Authors:  Hemal Bhatt; Mohammed Siddiqui; Eric Judd; Suzanne Oparil; David Calhoun
Journal:  J Am Soc Hypertens       Date:  2016-03-30

3.  Brachial-to-radial systolic blood pressure amplification in patients with type 2 diabetes mellitus.

Authors:  R E D Climie; D S Picone; M A Keske; J E Sharman
Journal:  J Hum Hypertens       Date:  2015-10-08       Impact factor: 3.012

4.  Accuracy in Blood Pressure Monitoring: The Effect of Noninvasive Blood Pressure Cuff Inflation on Intra-arterial Blood Pressure Values.

Authors:  Veena Sheshadri; Akhilesh Kumar Tiwari; Mahesh Nagappa; Lashmi Venkatraghavan
Journal:  Anesth Essays Res       Date:  2017 Jan-Mar

5.  Association of simultaneously measured four-limb blood pressures with cardiovascular function: a cross-sectional study.

Authors:  Xiaorui Song; Gaoyang Li; Aike Qiao; Zhihui Chen
Journal:  Biomed Eng Online       Date:  2016-12-28       Impact factor: 2.819

6.  Two-Year Responses of Office and Ambulatory Blood Pressure to First Occupational Lead Exposure.

Authors:  Yu-Ling Yu; Wen-Yi Yang; Lutgarde Thijs; Jesus D Melgarejo; Cai-Guo Yu; Dong-Mei Wei; Fang-Fei Wei; Tim S Nawrot; Zhen-Yu Zhang; Jan A Staessen
Journal:  Hypertension       Date:  2020-09-09       Impact factor: 10.190

7.  The cuffless SOMNOtouch NIBP device shows poor agreement with a validated oscillometric device during 24-h ambulatory blood pressure monitoring.

Authors:  Jakob Nyvad; Kent L Christensen; Niels Henrik Buus; Mark Reinhard
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-12-17       Impact factor: 3.738

8.  Effect of self-initiated and fully-automated self-measurement on blood pressure.

Authors:  Rosanne T Berkhof; Katia Gazzola; Bert-Jan H van den Born
Journal:  J Hum Hypertens       Date:  2019-09-23       Impact factor: 3.012

9.  Simultaneous compared with sequential blood pressure measurement results in smaller inter-arm blood pressure differences.

Authors:  Niels V van der Hoeven; Sophie Lodestijn; Stephanie Nanninga; Gert A van Montfrans; Bert-Jan H van den Born
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-09-19       Impact factor: 3.738

10.  Could the arm blood pressure measured with simultaneous bilateral arm method be used for hypertension diagnosis?

Authors:  Taixuan Wan; Yuanhao Wu; Ziqiang He; Hai Su
Journal:  BMJ Open       Date:  2020-09-16       Impact factor: 2.692

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.