Prospective randomized trial of wavefront-guided laser in situ keratomileusis with the CustomCornea and CustomVue laser systems.

Majid Moshirfar1, Ladan Espandar, Jay J Meyer, Justin R Tanner, Huck A Holz.   

Abstract

PURPOSE: To compare visual function, safety, and higher-order aberrations (HOAs) after wavefront-guided laser in situ keratomileusis (LASIK) with the LadarVision CustomCornea (Alcon Laboratories, Inc.) and Star S4 CustomVue (Visx) laser systems.
SETTING: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
METHODS: Seventy-eight eyes of 39 patients with myopia with or without astigmatism were randomized for LASIK treatment in 1 eye with the CustomCornea laser; the other eye was treated with the CustomVue laser. Patients were followed for 6 months after surgery. The primary outcome measures were uncorrected visual acuity (UCVA), best spectacle-corrected visual acuity (BSCVA), manifest refraction, and changes in HOAs.
RESULTS: At 6 months, the mean logarithm of the minimum angle of resolution (logMAR) UCVA was -0.0135 +/- 0.07 (SD) in the CustomVue group and 0.0417 +/- 0.12 in the CustomCornea group (P = .023). Eighty-eight percent of eyes in the CustomVue group had 20/20 or better UCVA compared with 67% in the CustomCornea group (P<.02). At 6 months, 91% of eyes in the CustomVue group and 79% in the CustomCornea group were within +/-0.50 diopter (D) of emmetropia (P<.1); 88% and 50%, respectively, were within +/-0.25 D (P<.001). Both platforms led to a small increase in total HOAs. The CustomVue system reduced trefoil and induced less of an increase in total HOAs, whereas the CustomCornea platform increased trefoil but induced less of an increase in spherical aberrations and coma.
CONCLUSIONS: Both laser systems were effective, safe, and predictable. Wavefront-guided LASIK with the CustomVue system resulted in better visual acuity, with more eyes having 20/20 acuity than in the CustomCornea group.

Entities:  

Mesh:

Year:  2007        PMID: 17889767     DOI: 10.1016/j.jcrs.2007.06.037

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  The association between femtosecond laser flap parameters and ocular aberrations after uncomplicated custom myopic LASIK.

Authors:  Christopher T Hood; Ronald R Krueger; Steven E Wilson
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-04-09       Impact factor: 3.117

2.  Wavefront excimer laser refractive surgery for adults with refractive errors.

Authors:  Shi-Ming Li; Meng-Tian Kang; Ning-Li Wang; Samuel A Abariga
Journal:  Cochrane Database Syst Rev       Date:  2020-12-18

3.  A prospective, randomized, fellow eye comparison of WaveLight® Allegretto Wave ® Eye-Q versus VISX CustomVue™ STAR S4 IR™ in laser in situ keratomileusis (LASIK): analysis of visual outcomes and higher order aberrations.

Authors:  Majid Moshirfar; Brent S Betts; Daniel S Churgin; Maylon Hsu; Marcus Neuffer; Shameema Sikder; Dane Church; Mark D Mifflin
Journal:  Clin Ophthalmol       Date:  2011-09-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.