Literature DB >> 17884055

Targeting device for intramedullary nails: a new high-stable mechanical guide.

Y Arlettaz1, A Akiki, F Chevalley, P F Leyvraz.   

Abstract

One of the most demanding steps of an intramedullary nailing is the distal locking. The aim of this study is to evaluate clinically and prospectively a new targeting guide. Twenty-five patients were treated consecutively in two different trauma centres (level one and one general hospital) by two different surgeons who were instructed on using the new device. Reduction of the fracture, intramedullary nailing and proximal locking are performed according to standard procedures. Then the new targeting device is used for distal locking. Evaluations are focused on the time it takes to perform the distal locking, the duration of the irradiation exposure and the duration of the complete procedure. In terms of duration, our results are comparable to those reported in the literature but this new device has more advantages than any other system. This new device is fully mechanical and is solidly linked to the patient. It can be used with any existing radioscopic equipment in any hospital. The guide is manually adjustable out of the X-ray field avoiding the surgeon being irradiated. The procedure is easy to learn and reproducible.

Entities:  

Mesh:

Year:  2007        PMID: 17884055     DOI: 10.1016/j.injury.2007.06.006

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  5 in total

1.  [Suprapatellar approach to tibial medullary nailing with electromagnetic field-guided distal locking].

Authors:  J M Rueger; A H Rücker; M Hoffmann
Journal:  Unfallchirurg       Date:  2015-04       Impact factor: 1.000

Review 2.  Meta-analysis suggests that the electromagnetic technique is better than the free-hand method for the distal locking during intramedullary nailing procedures.

Authors:  Yanbin Zhu; Hengrui Chang; Yiyang Yu; Wei Chen; Song Liu; Yingze Zhang
Journal:  Int Orthop       Date:  2016-06-02       Impact factor: 3.075

3.  Fluoroscopic freehand and electromagnetic-guided targeting system for distal locking screws of humeral intramedullary nail.

Authors:  P Persiani; M Gurzi; O Moreschini; G Di Giacomo; C Villani
Journal:  Musculoskelet Surg       Date:  2016-11-22

4.  Is a magnetic-manual targeting device an appealing alternative for distal locking of tibial intramedullary nails?

Authors:  Lukas L Negrin; Vilmos Vécsei
Journal:  Arch Trauma Res       Date:  2013-06-01

5.  Intramedullary nail holes laser indicator, a non-invasive technique for interlocking of intramedullary nails.

Authors:  Mohammadreza Maleki; Alireza Fadaei Tehrani; Ayda Aray; Mehdi Ranjbar
Journal:  Sci Rep       Date:  2021-10-27       Impact factor: 4.379

  5 in total