Spatial dependence of MLC transmission in IMRT delivery.

Friedlieb Lorenz1, Adrian Nalichowski, Florin Rosca, Jong Kung, Frederik Wenz, Piotr Zygmanski.   

Abstract

In complex intensity-modulated radiation therapy cases, a considerable amount of the total dose may be delivered through closed leaves. In such cases an accurate knowledge of spatial characteristics of multileaf collimator (MLC) transmission is crucial, especially for the treatment of large targets with split fields. Measurements with an ionization chamber, radiographic films (EDR2, EBT) and EPID are taken to characterize all relevant effects related to MLC transmission for various field sizes and depths. Here we present a phenomenological model to describe MLC transmission, whereby the main focus is the off-axis decrease of transmission for symmetric and asymmetric fields as well as on effects due to the tongue and groove design of the leaves, such as interleaf transmission and the tongue and groove effect. Data obtained with the four different methods are presented, and the utility of each measurement method to determine the necessary model parameters is discussed. With the developed model, it is possible to predict the relevant MLC effects at any point in the phantom for arbitrary jaw settings and depths.

Entities:  

Mesh:

Year:  2007        PMID: 17881814     DOI: 10.1088/0031-9155/52/19/018

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Validation of Dosimetric Leaf Gap (DLG) prior to its implementation in Treatment Planning System (TPS): TrueBeam™ millennium 120 leaf MLC.

Authors:  Ravindra Shende; Ganesh Patel
Journal:  Rep Pract Oncol Radiother       Date:  2017-10-21

Review 2.  Evaluation of fluence-based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG).

Authors:  Lalith K Kumaraswamy; Zhengzheng Xu; Daniel W Bailey; Jonathan D Schmitt; Matthew B Podgorsak
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

3.  Calculation of exit dose for conformal and dynamically-wedged fields, based on water-equivalent path length measured with an amorphous silicon electronic portal imaging device.

Authors:  Awusi Kavuma; Martin Glegg; Mohamed Metwaly; Garry Currie; Alex Elliott
Journal:  J Appl Clin Med Phys       Date:  2011-03-03       Impact factor: 2.102

4.  Static MLC transmission simulation using two-dimensional ray tracing.

Authors:  David P Adam; Bryan P Bednarz; Sean P Frigo
Journal:  J Appl Clin Med Phys       Date:  2022-05-20       Impact factor: 2.243

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.