The value of EDR2 film dosimetry in compensator-based intensity modulated radiation therapy.

R P Srivastava1, C De Wagter.   

Abstract

Radiographic or silver halide film is a well-established 2D dosimeter with an unquestioned spatial resolution. But its higher sensitivity to low-energy photons has to be taken into consideration. Metal compensators or physical modulators to deliver intensity modulated radiation therapy (IMRT) are known to change the beam energy spectrum and to produce scattered photons and contaminating electrons. Therefore the reliability of film dosimetry in compensator-based IMRT might be questioned. Conflicting data have been reported in the literature. This uncertainty about the validity of film dosimetry in compensator-based IMRT triggered us to conduct this study. First, the effect of MCP-96 compensators of varying thickness on the depth dose characteristics was investigated using a diamond detector which has a uniform energy response. A beam hardening effect was observed at 6 MV that resulted in a depth dose increase that remained below 2% at 20 cm depth. At 25 MV, in contrast, beam softening produced a dose decrease of up to 5% at the same depth. Second, dose was measured at depth using EDR2 film in perpendicular orientation to both 6 MV and 25 MV beams for different compensator thicknesses. A film dose underresponse of 1.1% was found for a 30 mm thick block in a 25 MV beam, which realized a transmission factor of 0.243. The effect induced by the compensators is higher than the experimental error but still within the accepted overall uncertainty of film dosimetry in clinical IMRT QA. With radiographic film as an affordable QA tool, the physical compensator remains a low threshold technique to deliver IMRT.

Entities:  

Mesh:

Year:  2007        PMID: 17881795     DOI: 10.1088/0031-9155/52/19/N03

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Simulation for improvement of system sensitivity of radiochromic film dosimetry with different band-pass filters and scanner light intensities.

Authors:  Takeshi Kamomae; Yuki Miyabe; Akira Sawada; Osamu Matoba; Manabu Nakata; Shinsuke Yano; Toru Takakura; Takashi Mizowaki; Akio Itoh; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2011-03-17

2.  Dosimetric verification of compensated beams using radiographic film.

Authors:  Slaven Jurkovic; Gordana Zauhar; Dario Faj; Deni Smilovic Radojcic; Manda Svabic; Mladen Kasabasic; Ana Diklic
Journal:  Radiol Oncol       Date:  2011-07-20       Impact factor: 2.991

3.  Evaluation of dose distribution of an intraoperative radiotherapy device using thermoluminescent dosimeters and radiographic films.

Authors:  Omid Baziar; Hamid Gholamhosseinian; Seyed Amir Aledavood; Mahboubeh Sadeghi
Journal:  Rep Pract Oncol Radiother       Date:  2022-07-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.