INTRODUCTION: This study reports the results of the clinical use of a new tourniquet system for surgery of the lower extremity that can determine tourniquet pressure in synchrony with systolic blood pressure.
MATERIALS AND METHODS: We prospectively applied additional pressure of 100 mmHg during 100 surgical procedures (54 knees, 2 legs, 44 feet and ankles).
RESULTS: A bloodless surgical field was obtained in almost all patients. The average durations of surgery and tourniquet time were 50 and 51 min, respectively. The average initial blood pressures were 112 mmHg systolic and 60 mmHg diastolic. The average maximum pressure changes during surgery were 33 mmHg systolic and 21 mmHg diastolic. No complications associated with this system arose either during or after surgery.
CONCLUSION: The conventional tourniquet pressure in the lower extremity is 300 or 350 mmHg, whereas this synchronized system required less pressure, making it a safe, useful device for controlling a bloodless surgical field in lower extremity surgery.

Entities:  

Mesh:

Year:  2007        PMID: 17874247     DOI: 10.1007/s00402-007-0446-0

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  3 in total

1.  Development of adaptive pneumatic tourniquet systems based on minimal inflation pressure for upper limb surgeries.

Authors:  Hong-yun Liu; Jun-yan Guo; Zheng-bo Zhang; Kai-yuan Li; Wei-dong Wang
Journal:  Biomed Eng Online       Date:  2013-09-23       Impact factor: 2.819

Review 2.  Tourniquet application during anesthesia: "What we need to know?"

Authors:  Kamal Kumar; Craig Railton; Qutaiba Tawfic
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2016 Oct-Dec

3.  Efficacy and safety of a new elastic tourniquet cuff in total knee arthroplasty: a prospective randomized controlled study.

Authors:  O-Sung Lee; Myung Chul Lee; Hyuk-Soo Han
Journal:  Biomed Eng Online       Date:  2017-08-08       Impact factor: 2.819

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.