Performance of heated wire humidifiers: an in vitro study.

Paolo Pelosi1, Davide Chiumello, Paolo Severgnini, Cinzia Elisabetta De Grandis, Laura Landi, Laura Maria Chierichetti, Alice Frigerio, Marco Munaro, Gilbert Park, Maurizio Chiaranda.   

Abstract

PURPOSE: We assessed the performance of heated wire humidifiers (HWHs), which should avoid water condensation in the circuit.
METHODS: We evaluated the efficiency of 3 HWHs, MR850 (Fisher & Paykel, Auckland, New Zealand), CONCHATHERM IV (Hudson RCI, Temecula, Calif), and DAR HC 2000 (Mallinckrodt DAR, Mirandola, Italy), in comparison with that of the MR730 heated humidifier (HH), which has a standard circuit. We measured gas temperature and absolute humidity (AH) at the Y piece of the ventilatory circuit using a test lung ventilated at 2 minute ventilation volumes (5 and 15 L/min). Temperature levels at the Y piece of the ventilatory circuit of the HHs were set at 35 degrees C, 37 degrees C, and 39 degrees C with different gradients (-2 degrees C, 0 degrees C, and +2 degrees C) between the outlet chamber and the Y piece of the ventilatory circuit.
RESULTS: At the set temperature levels of 35 degrees C, 37 degrees C, and 39 degrees C with a gradient of 0 degrees C, the MR850 and CONCHATHERM IV had lower gas temperature and AH levels as compared with the DAR HC 2000 and MR730 HH. With increasing temperature gradient, gas temperature increased only with the CONCHATHERM IV but AH increased with all the HWHs. The MR850 showed lower gas temperature and AH levels as compared with CONCHATHERM IV. The condensate was abolished inside the inspiratory circuit with the HWHs.
CONCLUSIONS: Heated wire humidifiers eliminate water condensation but present significant differences in gas temperature and AH levels that are lower than the expected settings.

Entities:  

Mesh:

Year:  2007        PMID: 17869979     DOI: 10.1016/j.jcrc.2006.12.009

Source DB:  PubMed          Journal:  J Crit Care        ISSN: 0883-9441            Impact factor:   3.425


  2 in total

1.  Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet.

Authors:  Davide Chiumello; Monica Chierichetti; Federica Tallarini; Paola Cozzi; Massimo Cressoni; Federico Polli; Riccardo Colombo; Antonio Castelli; Luciano Gattinoni
Journal:  Crit Care       Date:  2008-04-21       Impact factor: 9.097

2.  Insufficient Humidification of Respiratory Gases in Patients Who Are Undergoing Therapeutic Hypothermia at a Paediatric and Adult Intensive Care Unit.

Authors:  Yukari Tanaka; Sachiko Iwata; Masahiro Kinoshita; Kennosuke Tsuda; Shoichiro Tanaka; Naoko Hara; Ryota Shindou; Eimei Harada; Ryouji Kijima; Osamu Yamaga; Hitoe Ohkuma; Kazuo Ushijima; Teruo Sakamoto; Yushiro Yamashita; Osuke Iwata
Journal:  Can Respir J       Date:  2017-04-20       Impact factor: 2.409

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.