Mike Crilly1, Christoph Coch, Margaret Bruce, Hazel Clark, David Williams.
Abstract
Pulse wave analysis (PWA) using applanation tonometry is a non-invasive technique for assessing cardiovascular function. It produces three important indices: ejection duration index (ED%), augmentation index adjusted for heart rate (AIX@75), and subendocardial viability ratio (SEVR%). The aim of this study was to assess within- and between-observer repeatability of these measurements. After resting supine for 15 minutes, 20 ambulant patients (16 male) in sinus rhythm underwent four PWA measurements on a single occasion. Two nurses (A & B) independently and alternately undertook PWA measurements using the same equipment (Omron HEM-757; SphygmoCor with Millar hand-held tonometer) blind to the other nurse's PWA measurements. Within- and between-observer differences were analysed using the Bland-Altman ;limits of agreement' approach (mean difference +/- 2 standard deviations, 2SD). Mean age was 56 (blood pressure, BP 136/79; pulse rate 64). BP/PWA measurements remained stable during assessment. Based on the average of two PWA measurements the mean +/- 2SD between-observer difference in ED% was 0.3 +/- 2.0; AIX@75 1.0 +/- 3.9; and SEVR% 1.7 +/- 14.2. Based on a single PWA measurement the between-observer difference was ED% 0.3 +/- 3.3; AIX@75 1.7 +/- 6.9; and SEVR% 0.6 +/- 22.6. Within-observer differences for nurse-A were ED% 0.0 +/- 5.4; AIX@75 1.5 +/- 7.0; and SEVR% 1.7 +/- 39.0 (nurse-B: 0.1 +/- 3.8; 0.1 +/- 8.0; and 0.6 +/- 23.3, respectively). PWA demonstrates high levels of repeatability even when used by relatively inexperienced staff and has the potential to be included in the routine cardiovascular assessment of ambulant patients.Entities:
Mesh:
Year: 2007 PMID: 17848475 DOI: 10.1177/1358863X07081134
Source DB: PubMed Journal: Vasc Med ISSN: 1358-863X Impact factor: 3.239