Literature DB >> 17826286

Catastrophic wear in a metal-on-ceramic total hip arthroplasty.

Juan Ramon Valentí1, Javier Del Río, Santiago Amillo.   

Abstract

A 51-year-old woman came to our clinic 6 months after a right total hip arthroplasty. She had noticed a slowly growing mass in the proximal thigh and referred progressive pain in the right groin. Plain radiography revealed premature acetabular cup aseptic loosening, and in the computed tomography study, a 14-cm-diameter mass was observed. Revision surgery was performed, showing a metal-on-ceramic bearing surface. The histologic analysis of surrounding tissues was reported as massive metallosis. Although occasionally chosen for primary or revision hip arthroplasty, there is little information available about the in vivo wear behavior of this combination. This important fact should be taken into account before considering such a surface alternative.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17826286     DOI: 10.1016/j.arth.2006.06.008

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  5 in total

1.  Wear performance of ceramic-on-metal hip bearings.

Authors:  Jörn Reinders; Robert Sonntag; Christian Heisel; Tobias Reiner; Leo Vot; Jan Philippe Kretzer
Journal:  PLoS One       Date:  2013-08-29       Impact factor: 3.240

2.  Adverse reaction to metal debris in a patient with acetabular shell loosening 8 years after ceramic-on-metal total hip arthroplasty.

Authors:  Benjamin R Pulley; Thai Q Trinh; Jared C Bentley; Joel R Politi
Journal:  Arthroplast Today       Date:  2015-10-21

3.  Pseudotumor in ceramic-on-metal total hip arthroplasty.

Authors:  Yoni M Blau; Andrew J Meyers; Mauro Giordani; John P Meehan
Journal:  Arthroplast Today       Date:  2017-10-06

4.  Pseudotumour Complicated by Implant Loosening One Year After revision Ceramic on Metal Total Hip Arthroplasty: A Case Report.

Authors:  Devendra Lakhotia; Prashant Kumar
Journal:  J Orthop Case Rep       Date:  2017 Jan-Feb

5.  Ceramic-on-metal coupling in THA: long term clinical and radiographic outcomes using two different short stems.

Authors:  Giandomenico Logroscino; Michela Saracco; Giulio Maccauro; Andrea Urbani; Domenico Ciavardelli; Ada Consalvo; Daniele Ferraro; Francesco Falez
Journal:  BMC Musculoskelet Disord       Date:  2022-02-22       Impact factor: 2.362

  5 in total